Čolić, K. - Sedmak, A. - Legweel, K. - Milošević, M.I. - Nitrović, N. - Mišković, Ž. - Hloch, Sergej
Experimental and numerical research of mechanical behaviour of titanium alloy hip implant.
Technicki vjesnik - Technical Gazette. Roč. 24, č. 3 (2017), s. 709-713. ISSN 1330-3651. E-ISSN 1848-6339
Institucionální podpora: RVO:68145535
Klíčová slova: biomedical implant * digital image correlation * finite element method * Ti-6Al-4V alloy
Obor OECD: Mechanical engineering
Impakt faktor: 0.686, rok: 2017
https://hrcak.srce.hr/index.php?id_clanak_jezik=269849&show=clanak
Experimental and numerical research of mechanical behaviour of hip replacement implant is presented. Experimental analysis was based on standard methods for testing mechanical properties and application of Digital Image Correlation (DIC) technique, while the Finite Element Method (FEM) was used for numerical simulation. A hip prosthesis extracted from a patient after revision surgery was used as the specimen. An adapter for the implant testing was constructed to enable the condition of a proximally loosened stem. During FEM analysis, a solid model was made of the assembly. An analysis of presented results indicated good agreement of experimental and numerical research.
Trvalý link: http://hdl.handle.net/11104/0277242