Experimental and numerical research of mechanical behaviour of titanium alloy hip implant

  1. 1. 0481735 - UGN-S 2018 RIV HR eng J - Článek v odborném periodiku
    Čolić, K. - Sedmak, A. - Legweel, K. - Milošević, M.I. - Nitrović, N. - Mišković, Ž. - Hloch, Sergej
    Experimental and numerical research of mechanical behaviour of titanium alloy hip implant.
    Technicki vjesnik - Technical Gazette. Roč. 24, č. 3 (2017), s. 709-713 ISSN 1330-3651
    Institucionální podpora: RVO:68145535
    Klíčová slova: biomedical implant * digital image correlation * finite element method * Ti-6Al-4V alloy
    Kód oboru RIV: JQ - Strojní zařízení a nástroje
    Obor OECD: Mechanical engineering
    Impakt faktor: 0.686, rok: 2017
    https://hrcak.srce.hr/index.php?id_clanak_jezik=269849&show=clanak

    Experimental and numerical research of mechanical behaviour of hip replacement implant is presented. Experimental analysis was based on standard methods for testing mechanical properties and application of Digital Image Correlation (DIC) technique, while the Finite Element Method (FEM) was used for numerical simulation. A hip prosthesis extracted from a patient after revision surgery was used as the specimen. An adapter for the implant testing was constructed to enable the condition of a proximally loosened stem. During FEM analysis, a solid model was made of the assembly. An analysis of presented results indicated good agreement of experimental and numerical research.
    Trvalý link: http://hdl.handle.net/11104/0277242
    Název souboruStaženoVelikostKomentářVerzePřístup
    UGN_0481735.pdf02.2 MBVydavatelský postprintpovolen