Počet záznamů: 1
Calibration of windowless photodiode for extreme ultraviolet pulse energy measurement
- 1.0456145 - ÚFP 2016 RIV US eng J - Článek v odborném periodiku
Koláček, Karel - Schmidt, Jiří - Štraus, Jaroslav - Frolov, Oleksandr
Calibration of windowless photodiode for extreme ultraviolet pulse energy measurement.
Applied Optics. Roč. 54, č. 35 (2015), s. 10454-10459. ISSN 1559-128X. E-ISSN 2155-3165
Grant CEP: GA MŠMT(CZ) LG13029
Institucionální podpora: RVO:61389021
Klíčová slova: Photodetectors * Soft-X-ray * Extreme ultraviolet * Detection * Filters * Metrology
Kód oboru RIV: BH - Optika, masery a lasery
Impakt faktor: 1.598, rok: 2015
For energy measurement of extreme ultraviolet pulses, no universal commercially available device is available. Therefore, a co-axial setup of a vacuum photodiode was developed and tested. First its unsuccessful calibration at the National Institute of Standards and Technology (USA) is mentioned. Then our own single-wavelength (46.9 nm) calibration procedure is described in detail. It is based on tabulated photoemission efficiency and on measured (1) anode collection efficiency, (2) attenuation of Al filters, and (3) attenuation of measuring cables. Some other relevant attenuating factors are also taken into account.
Trvalý link: http://hdl.handle.net/11104/0256711
Počet záznamů: 1