Počet záznamů: 1
Integrated liquid chromatography–heated nebulizer microchip for mass spectrometry
- 1.0354131 - MBÚ 2011 RIV NL eng J - Článek v odborném periodiku
Haapala, M. - Saarela, V. - Pól, Jaroslav - Kolari, K. - Kotiaho, T. - Franssila, S. - Kostiainen, R.
Integrated liquid chromatography–heated nebulizer microchip for mass spectrometry.
Analytica Chimica Acta. Roč. 662, č. 2 (2010), s. 163-169. ISSN 0003-2670. E-ISSN 1873-4324
Výzkumný záměr: CEZ:AV0Z50200510
Klíčová slova: Liquid chromatography * Mass spectrometry * Microchip
Kód oboru RIV: CB - Analytická chemie, separace
Impakt faktor: 4.310, rok: 2010
A new integrated microchip for liquid chromatography–mass spectrometry (LC–MS) is presented. The chip is made from bonded silicon and glass wafers with structures for a packed LC column channel, a micropillar frit, a channel for optional optical detection, and a heated vaporizer section etched in silicon and platinum heater elements on the glass cover. LC eluent is vaporized and mixed with nebulizer gas in the vaporizer section and the vapor is sprayed out from the chip. Nonpolar and polar analytes can be efficiently ionized in the gas phase by atmospheric pressure photoionization (APPI) as demonstrated with polycyclic aromatic hydrocarbons (PAHs) and selective androgen receptor modulators (SARMs). This is not achievable with present LC–MS chips, since they are based on electrospray ionization, which is not able to ionize nonpolar compounds efficiently
Trvalý link: http://hdl.handle.net/11104/0193201
Počet záznamů: 1