Počet záznamů: 1
Radial flow enhances QCM biosensor sensitivity
- 1.
SYSNO ASEP 0585171 Druh ASEP J - Článek v odborném periodiku Zařazení RIV J - Článek v odborném periodiku Poddruh J Článek ve WOS Název Radial flow enhances QCM biosensor sensitivity Tvůrce(i) Lynn, Nicholas Scott (FZU-D) ORCID
Forinová, Michala (FZU-D) ORCID
Spasovová, Monika (FZU-D) ORCID
Vaisocherová-Lísalová, Hana (FZU-D) ORCID
Yudin, Petr (FZU-D) ORCIDCelkový počet autorů 5 Číslo článku 134949 Zdroj.dok. Sensors and Actuators B - Chemical. - : Elsevier
Roč. 401, Feb (2024)Poč.str. 8 s. Jazyk dok. eng - angličtina Země vyd. NL - Nizozemsko Klíč. slova biosensor ; quartz crystal microbalance ; radial flow ; microfluidics ; diffusion limitations ; detection of Escherichia coli Vědní obor RIV BO - Biofyzika Obor OECD Biophysics CEP GA22-20012S GA ČR - Grantová agentura ČR Institucionální podpora FZU-D - RVO:68378271 UT WOS 001146836700001 EID SCOPUS 85177204100 DOI 10.1016/j.snb.2023.134949 Anotace QCM biosensors are widely used for the detection of a variety of pathogenic microorganisms, and there remains a need for further increases in sensitivity. The majority of QCM biosensors are operated under diffusion-limited conditions that, when used with a standard flow cell, lead to the majority of analyte being captured onto re- gions of the crystal surface that have limited to no mass sensitivity. Here we show that simple changes to a flow cell architecture (reduction of flow cell height) can provide increases in sensing signal. Furthermore, we show that the use of radial flow will lead to analyte capture directly in the sensitive region of a crystal, thus providing dramatic increases in the biosensor’s sensitivity. We examine this problem from both a theoretical and numerical perspective and present a new method by which to consider how changes to a sensing assay will affect sensor sensitivity. Pracoviště Fyzikální ústav Kontakt Kristina Potocká, potocka@fzu.cz, Tel.: 220 318 579 Rok sběru 2025
Počet záznamů: 1