Počet záznamů: 1  

Round robin comparison study on the form measurement of optical freeform surfaces

  1. 1.
    SYSNO ASEP0541106
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevRound robin comparison study on the form measurement of optical freeform surfaces
    Tvůrce(i) Fortmeier, I. (DE)
    Schachtschneider, R. (DE)
    Lédl, Vít (UFP-V) RID
    Matoušek, Ondřej (UFP-V)
    Siepmann, J. (DE)
    Harsch, A. (DE)
    Beisswanger, R. (DE)
    Bitou, Y. (JP)
    Kondo, Y. (JP)
    Schulz, M. (DE)
    Elster, C. (DE)
    Celkový počet autorů11
    Číslo článku2
    Zdroj.dok.Journal of the European Optical Society: Rapid publications. - : European optical society
    Roč. 16, č. 1 (2020)
    Poč.str.15 s.
    Jazyk dok.eng - angličtina
    Země vyd.GB - Velká Británie
    Klíč. slovaFreeform optical surfaces ; Interlaboratory comparison ; Metrology
    Vědní obor RIVBH - Optika, masery a lasery
    Obor OECDOptics (including laser optics and quantum optics)
    Způsob publikováníOpen access
    Institucionální podporaUFP-V - RVO:61389021
    UT WOS000511497800001
    EID SCOPUS85077508507
    DOI10.1186/s41476-019-0124-1
    AnotaceA round robin comparison of freeform form measurements was carried out by the project partners and stakeholders of a European metrology research project. Altogether six measuring instruments were considered: five different (pointwise and areal) optical devices and one tactile device. Three optical freeform surfaces were used for the comparison measurements, where two specimens were measured by five instruments and one specimen by four instruments. In this paper, the evaluation methods and results of this round robin are presented for the three freeform surfaces made from a temperature-stable material, Super Invar ®. The freeforms had diameters of 40 mm, 50 mm and 100 mm and best-fit radii of 39.75 mm (convex), 40.9 mm (convex) and 423.5 mm (concave). For comparison, the bilateral pointwise differences between the available measurements were calculated. The root-mean-square values of these differences ranged from 15 nm to 110 nm (neglecting spherical contributions) and provided an insight into the status of typical freeform measurement capabilities for optical surfaces.
    PracovištěÚstav fyziky plazmatu
    KontaktVladimíra Kebza, kebza@ipp.cas.cz, Tel.: 266 052 975
    Rok sběru2021
    Elektronická adresahttps://jeos.springeropen.com/track/pdf/10.1186/s41476-019-0124-1.pdf