Počet záznamů: 1  

Separation of anaesthetic ketamine and its derivates in PAMAPTAC coated capillaries with tuneable counter-current electroosmotic flow

  1. 1.
    SYSNO ASEP0524392
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevSeparation of anaesthetic ketamine and its derivates in PAMAPTAC coated capillaries with tuneable counter-current electroosmotic flow
    Tvůrce(i) Tůma, P. (CZ)
    Koval, Dušan (UOCHB-X) RID, ORCID
    Sommerová, B. (CZ)
    Vaculín, Š. (CZ)
    Číslo článku121094
    Zdroj.dok.Talanta. - : Elsevier - ISSN 0039-9140
    Roč. 217, Sep (2020)
    Poč.str.8 s.
    Jazyk dok.eng - angličtina
    Země vyd.NL - Nizozemsko
    Klíč. slovaclinical analysis ; capillary electrophoresis ; coating ; contactless conductivity detection ; Peak master ; stacking
    Vědní obor RIVCA - Anorganická chemie
    Obor OECDAnalytical chemistry
    CEPGA18-02597S GA ČR - Grantová agentura ČR
    Způsob publikováníOmezený přístup
    Institucionální podporaUOCHB-X - RVO:61388963
    UT WOS000537880200084
    EID SCOPUS85084174760
    DOI10.1016/j.talanta.2020.121094
    AnotaceCapillary electrophoretic separation of ketamine, norketamine, hydroxynorketamine, and dehydronorketamine was performed in the counter-current regime under the influence of oppositely-directed electroosmotic flow. For this purpose, the fused silica capillaries were covalently coated with the poly(acrylamide-co-3-acrylamidopropyl trimethylammonium chloride) copolymer (PAMAPTAC). The content of the cationic monomer APTAC in the polymerization mixture varied in the range 0–6 mol. % and the generated electroosmotic flow increased continuously in the 0–20 · 10−9 m2V−1s−1 interval. Importantly, it resulted in improved electrophoretic resolution of ketamine/norketamine, which increased from 0.8 for neutral PAM coating (i.e. 0% PAMAPTAC) to 3.0 for 6% PAMAPTAC. The determination of ketamine and its derivates in rat serum was performed in a 4% PAMAPTAC capillary with an inner diameter of 25 μm. The separation was performed in a 500 mM aqueous solution of acetic acid (pH 2.3). The clinical sample was deproteinized by the addition of acetonitrile to the serum and a large volume of the treated sample was injected directly into the capillary. The achieved limit of detection ranged from 2.2 ng/mL for dehydronorketamine to 4.1 ng/mL for hydroxynorketamine, the intra-day repeatability was 1.0–1.5% for the migration time and 2.8–3.3% for the peak area. The developed methodology was employed for time monitoring of ketamines in rat serum after intra venous administration of low doses of anaesthetic at a level of 2 μg per g of body weight.
    PracovištěÚstav organické chemie a biochemie
    Kontaktasep@uochb.cas.cz ; Kateřina Šperková, Tel.: 232 002 584 ; Viktorie Chládková, Tel.: 232 002 434
    Rok sběru2021
    Elektronická adresahttps://www.sciencedirect.com/science/article/pii/S0039914020303854?via%3Dihub
Počet záznamů: 1  