Počet záznamů: 1  

Time-Dependent Oxidation during Nano-Assisted Laser Desorption Ionization Mass Spectrometry: A Useful Tool for Structure Determination or a Source of Possible Confusion?

  1. 1.
    SYSNO ASEP0369749
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevTime-Dependent Oxidation during Nano-Assisted Laser Desorption Ionization Mass Spectrometry: A Useful Tool for Structure Determination or a Source of Possible Confusion?
    Tvůrce(i) Pavlásková, Kateřina (MBU-M)
    Strnadová, Marcela (MBU-M)
    Strohalm, Martin (MBU-M)
    Havlíček, Vladimír (MBU-M) RID, ORCID
    Šulc, Miroslav (MBU-M) RID, ORCID
    Volný, Michael (MBU-M) ORCID
    Zdroj.dok.Analytical Chemistry. - : American Chemical Society - ISSN 0003-2700
    Roč. 83, č. 14 (2011), s. 5661-5665
    Poč.str.5 s.
    Jazyk dok.eng - angličtina
    Země vyd.US - Spojené státy americké
    Klíč. slovaDOUBLE-BOND POSITION ; UNSATURATED FATTY-ACIDS ; ELECTROSPRAY-IONIZATION
    Vědní obor RIVCE - Biochemie
    CEPGPP206/10/P018 GA ČR - Grantová agentura ČR
    LC545 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    ME10013 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    CEZAV0Z50200510 - MBU-M (2005-2011)
    UT WOS000292892000026
    DOI10.1021/ac200801t
    AnotaceThis work reports on a new and extremely simple approach for determination of a double bond position by a laser desorption ionization mass spectrometry. It is solely based on the catalytic properties of nanostructured surfaces used in nanoassisted laser desorption ionization experiments. These surfaces can induce oxidation of analytes, which results in a mass shift that can be detected by mass spectrometry. If a site of unsaturation is oxidized and cleaved, the m/z difference is diagnostic of the position of a double bond. By demonstrating that the oxidation depends on the analyte surface dwell time, it was proven that it is caused by the surface activity and not by the laser desorption ionization process itself. Control matrix-assisted laser desorption/ionization (MALDI)experiment showed only a limited partial oxidation and no time dependency of the process
    PracovištěMikrobiologický ústav
    KontaktEliška Spurná, eliska.spurna@biomed.cas.cz, Tel.: 241 062 231
    Rok sběru2012