Počet záznamů: 1  

Modeling and Measuring of the Changes of the Magnetic Field Next to Non-Ferromagnetic Substances for Magnetic Susceptibility Calculation

  1. 1.
    SYSNO ASEP0368262
    Druh ASEPC - Konferenční příspěvek (mezinárodní konf.)
    Zařazení RIVD - Článek ve sborníku
    NázevModeling and Measuring of the Changes of the Magnetic Field Next to Non-Ferromagnetic Substances for Magnetic Susceptibility Calculation
    Tvůrce(i) Marcon, P. (CZ)
    Bartušek, Karel (UPT-D) RID, ORCID, SAI
    Čáp, M. (CZ)
    Celkový počet autorů3
    Zdroj.dok.Measurement 2011. - Bratislava : Slovak Academy of Sciences, 2011 - ISBN 978-80-969672-4-7
    Rozsah strans. 213-216
    Poč.str.4 s.
    AkceInternational Conference on Measurement 2011 /8./
    Datum konání27.04.2011-30.04.2011
    Místo konáníSmolenice
    ZeměSK - Slovensko
    Typ akceEUR
    Jazyk dok.eng - angličtina
    Země vyd.SK - Slovensko
    Klíč. slovaCOMSOL ; non-ferromagnetic ; NMR ; magnetic susceptibility
    Vědní obor RIVJA - Elektronika a optoelektronika, elektrotechnika
    CEPGAP102/11/0318 GA ČR - Grantová agentura ČR
    ED0017/01/01 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    CEZAV0Z20650511 - UPT-D (2005-2011)
    AnotaceThis article is focused on the principles of the magnetic susceptibility measurement of the non-ferromagnetic substances using NMR tomography. Magnetic susceptibility is calculated from changes of the magnetic field close to the cylinder shaped specimen. COMSOL program was used for modeling experiment results. Consequently, model results were compared to MR tomograph measured data. Changes of the magnetic field in space of xy cut of the specimen were integrated in both cases. Before integrating of the magnetic field changes was necessary to filter, unwrap and detect accurate position of the specimen. Magnetic susceptibility calculated from measured data is in comparison to theoretical value, as well as model value, slightly different. Difference of those values is lower than 5%. Due to the used model and measurement were extended principles of the NMR method for magnetic susceptibility measurement.
    PracovištěÚstav přístrojové techniky
    KontaktMartina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178
    Rok sběru2012
Počet záznamů: 1  