SYSNO ASEP 0366067 Druh ASEP J - Článek v odborném periodiku Zařazení RIV J - Článek v odborném periodiku Poddruh J Článek ve WOS Název Magnetic susceptibility measurement using 2D magnetic resonance imaging Tvůrce(i) Marcon, P. (CZ)
Bartušek, Karel (UPT-D) RID, ORCID, SAI
Burdkova, M. (CZ)
Dokoupil, Zdeněk (UPT-D) RID, SAICelkový počet autorů 4 Zdroj.dok. Measurement Science and Technology. - : Institute of Physics Publishing - ISSN 0957-0233
Roč. 22, č. 10 (2011), 105702:1-8Poč.str. 8 s. Jazyk dok. eng - angličtina Země vyd. GB - Velká Británie Klíč. slova magnetic flux density ; magnetic susceptibility ; MRI ; MR signal ; reaction field Vědní obor RIV JA - Elektronika a optoelektronika, elektrotechnika CEP GAP102/11/0318 GA ČR - Grantová agentura ČR ED0017/01/01 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy CEZ AV0Z20650511 - UPT-D (2005-2011) UT WOS 000294819500045 EID SCOPUS 80052752276 DOI 10.1088/0957-0233/22/10/105702 Anotace The authors describe a nuclear magnetic resonance (NMR) method for measuring the magnetic susceptibility of bar-shaped samples that have an arbitrary cross-section and do not produce an MR signal. The method is based on the measurement of the 2D map of the reaction field in the vicinity of a non-ferromagnetic sample and on the calculation of magnetic susceptibility from a known reaction field. The verification of the technique was realized via modelling the measured configuration in the ANSYS program by means of the finite element method and through an experimental measurement of MR-compatible and MR-incompatible materials carried out on a tomograph. A great advantage of the proposed susceptibility evaluation method consists in the use of only standard commercially used devices without the need of any special sequences. Pracoviště Ústav přístrojové techniky Kontakt Martina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178 Rok sběru 2012
Počet záznamů: 1