Počet záznamů: 1  

Voltammetric determination of homovanillic acid at boron doped diamond electrode for hollow fibre microextraction

  1. 1.
    0470878 - ÚFCH JH 2017 RIV CZ eng C - Konferenční příspěvek (zahraniční konf.)
    Hrdlička, V. - Barek, J. - Navrátil, Tomáš
    Voltammetric determination of homovanillic acid at boron doped diamond electrode for hollow fibre microextraction.
    Proceedings of the 12th International students conference Modern Analytical Chemistry. Praha: VŠCHT, 2016 - (Nesměrák, K.), s. 206-210. ISBN 978-80-7444-044-1.
    [ISC Modern Analytical Chemistry /12./. Prague (CZ), 22.09.2016-23.09.2016]
    Institucionální podpora: RVO:61388955
    Klíčová slova: biomarkers * boron doped diamond electrode * hollow-fibre
    Kód oboru RIV: CG - Elektrochemie

    Presented work deals with hollow-fibre assisted liquid/liquid/liquid microextraction (HF-LPME) and subsequent voltammetric detection of clinical biomarker homovanillic acid (HVA). Three main areas are discussed: hollow fibre assisted extraction of HVA, voltammetric detection of HVA under basic conditions, and sensing in very small volumes. Using HF-PLME, HVA was extracted with 1-octanol as a supported liquid membrane with enrichment factor of 55 after 60 minutes of extraction. A novel method of miniaturized voltammetric determination of HVA using cathodically pretreated boron doped diamond electrode was also developed. Using differential pulse voltammetry, limit of detection, limit of quantificati was 1.9 mu mol dm(-3) on 6.6 mu mol dm(-3).
    Trvalý link: http://hdl.handle.net/11104/0268393

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.