Počet záznamů: 1
Analytical isotachophoresis of lactate in human serum using dry film photoresist microfluidic chips compatible with a commercially available field-deployable instrument platform
- 1.0398860 - ÚIACH 2014 RIV NL eng J - Článek v odborném periodiku
Smejkal, P. - Breadmore, M. C. - Guijt, R. M. - Foret, František - Bek, F. - Macka, M.
Analytical isotachophoresis of lactate in human serum using dry film photoresist microfluidic chips compatible with a commercially available field-deployable instrument platform.
Analytica Chimica Acta. Roč. 803, NOV (2013), s. 135-142. ISSN 0003-2670. E-ISSN 1873-4324
Grant CEP: GA ČR(CZ) GAP301/11/2055
Institucionální podpora: RVO:68081715
Klíčová slova: isotachophoresis * microfluidics * fluorescence * indirect detection * serum
Kód oboru RIV: CB - Analytická chemie, separace
Impakt faktor: 4.517, rok: 2013
A dry film resist (DFR) chip compatible with the Agilent Bioanalyzer 2100 was designed and fabricated for use in the analysis of lactate in serum by chip isotachophoresis (ITP). The Agilent Bioanalyzer 2100 is a commercially available field deployable analytical instrument originally developed for the electrophoretic analysis of DNA, RNA and proteins. The DFR chip was designed for the ITP separation of lactate in human serum within 1 min and was made compatible with the Bioanalyzer after packaging in the plastic caddies normally used for the DNA chips. A 20-fold improvement in sensitivity was obtained for the DFR chips in comparison with the standard chips used in earlier work. The limit of detection and limit of quantification for lactate were 24 M and 80 M, respectively. This new approach enables the use of commercial platforms like the Agilent Bioanalyzer for new applications including the analysis of small molecules.
Trvalý link: http://hdl.handle.net/11104/0226278
Název souboru Staženo Velikost Komentář Verze Přístup 0398860_IR.pdf 2 1.5 MB Jiná vyžádat