Počet záznamů: 1
Scintillation SE detector for variable pressure scanning electron microscopes
- 1.0350812 - ÚPT 2011 RIV GB eng J - Článek v odborném periodiku
Jirák, J. - Neděla, Vilém - Černoch, P. - Čudek, P. - Runštuk, Jiří
Scintillation SE detector for variable pressure scanning electron microscopes.
Journal of Microscopy. Roč. 239, č. 3 (2010), s. 233-238. ISSN 0022-2720. E-ISSN 1365-2818
Grant CEP: GA ČR GAP102/10/1410
Výzkumný záměr: CEZ:AV0Z20650511
Klíčová slova: scintillation detector * secondary electrons detection * variable pressure scanning electron
Kód oboru RIV: JA - Elektronika a optoelektronika, elektrotechnika
Impakt faktor: 1.872, rok: 2010
We present results obtained with a new scintillation detector of secondary electrons for the variable pressure scanning electron microscope. A detector design is based on the positioning of a single crystal scintillator within a scintillator chamber separated from the specimen chamber by two apertures. This solution enables us to decrease the pressure to several Pa in the scintillator chamber while the pressure in the specimen chamber reaches values of about 1000 Pa (7.5 Torr). Due to decreased pressure, we can apply a potential of the order of several kV to the scintillator, which is necessary for the detection of secondary electrons. Simultaneously, the two apertures at appropriate potentials of the order of several hundreds of volts create an electrostatic lens that allows electrons to pass from the specimen chamber to the scintillator chamber. Results indicate a promising utilization of this detector for a wide range of specimen observations.
Trvalý link: http://hdl.handle.net/11104/0190715
Počet záznamů: 1