Počet záznamů: 1
Hydride generation – in-atomizer collection of Pb in a quartz trap-and-atomizer device for atomic absorption spectrometry – an interference study
- 1.0388749 - ÚIACH 2013 RIV NL eng J - Článek v odborném periodiku
Novotný, Pavel - Kratzer, Jan
Hydride generation – in-atomizer collection of Pb in a quartz trap-and-atomizer device for atomic absorption spectrometry – an interference study.
Spectrochimica Acta Part B: Atomic Spectroscopy. 79-80, JAN-FEB (2013), s. 77-81. ISSN 0584-8547. E-ISSN 1873-3565
Grant CEP: GA ČR(CZ) GPP206/11/P002
Výzkumný záměr: CEZ:AV0Z40310501
Institucionální podpora: RVO:68081715
Klíčová slova: trap-and-atomizer device * lead determination * interference study
Kód oboru RIV: CB - Analytická chemie, separace
Impakt faktor: 3.150, rok: 2013 ; AIS: 0.76, rok: 2013
DOI: https://doi.org/10.1016/j.sab.2012.12.002
Interference study on Pb determination by in-atomizer trapping was performed, Bi was found as a severe interferent in the preconcentration mode. No interference was found in the on-line atomization (no preconcentration).
Trvalý link: http://hdl.handle.net/11104/0217664Název souboru Staženo Velikost Komentář Verze Přístup 0388749_IR.pdf 3 209.4 KB Jiná vyžádat
Počet záznamů: 1