Počet záznamů: 1  

Intravascular ultrasound study of the effect of beta-emitting (Co-55) stents on vascular remodeling and intimal proliferation

  1. 1.
    SYSNO ASEP0105596
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JOstatní články
    NázevIntravascular ultrasound study of the effect of beta-emitting (Co-55) stents on vascular remodeling and intimal proliferation
    Překlad názvuIntravaskulární echografická studie efektu beta-emitnujícího (55Co) stentu
    Tvůrce(i) Červinka, P. (CZ)
    Stašek, J. (CZ)
    Costa, MA. (BR)
    Štursa, Jan (UJF-V) RID, ORCID
    Fišer, Miroslav (UJF-V)
    Vodňanský, P. (CZ)
    Kocisová, M. (CZ)
    Veselka, J. (CZ)
    Pleskot, M. (CZ)
    Malý, J. (CZ)
    Zdroj.dok.Catheterization and Cardiovascular Interventions - ISSN 1522-1946
    Roč. 61, č. 3 (2004), s. 320-325
    Poč.str.26 s.
    Jazyk dok.eng - angličtina
    Země vyd.US - Spojené státy americké
    Klíč. slovaradiactive stents ; angioplasty ; remodeling
    Vědní obor RIVBG - Jaderná, atomová a mol. fyzika, urychlovače
    CEPKSK4055109 GA AV ČR - Akademie věd
    AnotaceThe aim of this study was to evaluate vessel remodeling after implantation of high-activity (mean, 41.1 +/- 1.2 muCi) beta-emitting ((CO)-C-55) stents. Proton bombarding in cyclotron has brought the radioactivity. Intravascular ultrasound (IVUS) investigation has been completed in 10 patients. The angiographies performed at 6 months revealed restenosis > 50% in five cases (50%). IVUS analysis demonstrated an absence of remodeling behind the stent, with no changes in total vessel volume (TVV; 353.6 +/- 126.3 and 343.9. +/- 90.6 mm(3)) or plaque + media volume (PMV; 171.7 +/- 57.4 and 166.8 +/- 42.6 mm(3)). On the other hand, lumen volume (LV) within the stent decreased significantly from 181.9 +/- 80.2 to 154.6 +/- 45.2mm(3) (P < 0.02). This was due to presence of neointimal hyperplasia (NIH) at both extremities of implanted stents. No chronic recoil of the implanted stents was found. The analysis of edges (5 mm distally and proximally to the last stent struts) showed no significant changes in TVV (187.3 &PLUSMN; 62.60 and 176.9 &PLUSMN; 53.5 mm(3)), but PMV increase significantly from 61.9 &PLUSMN; 31.2 to 82.2 &PLUSMN; 43.4 mm(3) (p < 0.04) and LV decreased from 125.2 +/- 40.7 to 94.7 +/- 22.0 mm(3) (p < 0.02). In conclusion, single Co-55 radioactive β-emitting stents with high initial activity are effective in reducing neointimal hyperplasia only within the stent body, as measured by IVUS, and they do not solve the problem of restenosis at the stent extremities as well as at the stent edges. Edge restenosis in this high radioactive stents was mainly (from 66%) due to neointimal proliferation.
    PracovištěÚstav jaderné fyziky
    KontaktMarkéta Sommerová, sommerova@ujf.cas.cz, Tel.: 266 173 228
    Rok sběru2005
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.