Počet záznamů: 1
Influence of stent implantation on wall compliance
- 1.0025551 - ÚT 2006 RIV IT eng C - Konferenční příspěvek (zahraniční konf.)
Horný, L. - Chlup, Hynek - Konvičková, S.
Influence of stent implantation on wall compliance.
[Vliv implantace stentu na poddajnost stěny trubice.]
22nd Danubia-Adria Symposium on Experimental Methods in Solid Mechanics. Parma, 2005, s. 1-2.
[Danubia-Adria Symposium on Experimental Methods in Solid Mechanics /22./ (28.9.2005-1.10.2005: Moticelli Terme). Moticelli Terme (IT), 28.09.2005-01.10.2005]
Grant CEP: GA ČR(CZ) GA106/04/1181
Výzkumný záměr: CEZ:AV0Z20760514
Klíčová slova: constitutive equations * coronary artery * stent
Kód oboru RIV: BO - Biofyzika
Inflation tests with latex tubes and human arteria coronaria dextra were performed. Compliance with and without implanted stent was measured by optical system and evaluated in terms of external diameter. Deformation characteristics of artery and latex tubes were compared.
Experimentálně byly určeny tlakově-deformační charakteristiky pro tepennou stěnu s implantovaným stentem a bez něj. Porovnávána byla poddajnost stěny. Zároveň byl srovnáván vliv implantace stentu do tepny s implantací do latexové trubice. Použitá tepna byla lidská pravá věnčitá tepna.
Trvalý link: http://hdl.handle.net/11104/0115928
Počet záznamů: 1