Relative efficiency of radiophotoluminescent glass dosimeters in a scanning pencil proton beam

  1. 1.
    SYSNO ASEP0581046
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevRelative efficiency of radiophotoluminescent glass dosimeters in a scanning pencil proton beam
    Tvůrce(i) Majer, M. (HR)
    Pasaricek, L. (HR)
    Brkic, H. (HR)
    Davídková, Marie (UJF-V) RID, ORCID, SAI
    Navrátil, M. (CZ)
    Vondráček, V. (CZ)
    Kneževic, Z. (HR)
    Celkový počet autorů7
    Číslo článku111396
    Zdroj.dok.Radiation Physics and Chemistry. - : Elsevier - ISSN 0969-806X
    Roč. 216, MAR (2024)
    Poč.str.9 s.
    Forma vydáníTištěná - P
    Jazyk dok.eng - angličtina
    Země vyd.GB - Velká Británie
    Klíč. slovaDose response ; Relative efficiency ; RPL glass dosimeter ; TLD-100 ; Proton therapy
    Obor OECDRadiology, nuclear medicine and medical imaging
    Způsob publikováníOmezený přístup
    Institucionální podporaUJF-V - RVO:61389005
    UT WOS001132601400001
    EID SCOPUS85178160954
    DOI10.1016/j.radphyschem.2023.111396
    AnotaceThe increasing role of hadron therapy, and particularly of proton therapy, in cancer treatment requires dosimetric characterization of luminescent detectors in heavy charged particle beams. The ionization density, which is related to the linear energy transfer (LET), is changing when heavy charged particle is penetrating through a material. Therefore, it is crucial to evaluate how efficiency of the used dosimeters depends on the change in LET.The aim of this study was to evaluate relative efficiency of radiophotoluminescent (RPL) glass dosimeters (type GD-352M) in a clinical scanning pencil proton beam of 100 MeV and compare it to the results obtained with thermoluminescent (TL) dosimeters (type TLD-100). LETf and LETd at the selected dosimeter's position were evaluated and compared by two Monte Carlo codes (MCNP and PHITS). Linear dose response is precondition for the evaluation of dosimeter efficiency and first, dose response in proton beam was evaluated and compared with those in 60Co gamma ray field.Linear dose response up to 9 Gy was confirmed for RPLs in proton beam, and gamma ray field. For TLDs, supralinearity from about 5 Gy was observed. A constant relative efficiency is observed along the plateau (corresponding to LETf from 0.7 keV/mu m to 1.7 keV/mu m) of non-modulated proton beam with an average value 1.00 +/- 0.02 for RPL and 0.97 +/- 0.02 for TL dosimeters. Relative efficiency decrease of 10 % is observed for RPLs within the Bragg peak with LETf = 2.3 keV/mu m. Future work will be focused on the evaluation of the efficiency for higher LET values using more energetic proton beams and precise positioning along the Bragg peak.