Počet záznamů: 1
Spatially and temporally resolved detection of arsenic in a capillary dielectric barrier discharge by hydride generation high-resolved optical emission spectrometry
- 1.0487597 - ÚIACH 2019 RIV US eng J - Článek v odborném periodiku
Burhenn, S. - Kratzer, Jan - Svoboda, Milan - Klute, F. D. - Michels, A. - Veža, D. - Franzke, J.
Spatially and temporally resolved detection of arsenic in a capillary dielectric barrier discharge by hydride generation high-resolved optical emission spectrometry.
Analytical Chemistry. Roč. 90, č. 5 (2018), s. 3424-3429. ISSN 0003-2700. E-ISSN 1520-6882
Grant CEP: GA ČR GA17-04329S
Institucionální podpora: RVO:68081715
Klíčová slova: dielectric barrier discharge * optical emission spectroscopy * arsenic hydride
Obor OECD: Analytical chemistry
Impakt faktor: 6.350, rok: 2018
An improved method for the detection of arsenic using a specially adapted setup for OES was presented. By comparing the temporal and spatial evolution of the arsenic emission signal with the emission of the dissociation of molecular hydrogen it was shown that produced hydrogen radicals play a major role in the atomization of arsine.
Trvalý link: http://hdl.handle.net/11104/0282247