Počet záznamů: 1  

Fatigue properties of UFG Ti grade 2 dental implant vs. conventionally tested smooth specimens

  1. 1.
    SYSNO ASEP0544713
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevFatigue properties of UFG Ti grade 2 dental implant vs. conventionally tested smooth specimens
    Tvůrce(i) Fintová, Stanislava (UFM-A) ORCID
    Dlhý, Pavol (UFM-A) ORCID
    Mertová, K. (CZ)
    Chlup, Zdeněk (UFM-A) RID, ORCID
    Duchek, M. (CZ)
    Procházka, R. (CZ)
    Hutař, Pavel (UFM-A) RID, ORCID
    Celkový počet autorů7
    Číslo článku104715
    Zdroj.dok.Journal of the Mechanical Behavior of Biomedical Materials. - : Elsevier - ISSN 1751-6161
    Roč. 123, NOV (2021)
    Poč.str.12 s.
    Jazyk dok.eng - angličtina
    Země vyd.NL - Nizozemsko
    Klíč. slovaDental implant ; Fatigue ; SLA treatment ; Ti grade 2 ; Ultrafine-grained
    Vědní obor RIVJG - Hutnictví, kovové materiály
    Obor OECDMaterials engineering
    Způsob publikováníOmezený přístup
    Institucionální podporaUFM-A - RVO:68081723
    UT WOS000703875400004
    EID SCOPUS85111842299
    DOI10.1016/j.jmbbm.2021.104715
    AnotaceComplicated geometry in combination with surface treatment strongly deteriorates fatigue resistance of metallic dental implants. Mechanical properties of pure Ti grade 2, usually used for dental implant production, were shown to be significantly improved due to intensive grain refinement via Conform SPD. The increase of the tensile strength properties was accompanied by a significant increase in the fatigue resistance and fatigue endurance limit. However, the SLA treatment usually used for the implants' surface roughening, resulted in the fatigue properties and endurance limit decrease, while this effect was more pronounced for the ultrafine-grained comparing to the coarse-grained material when tested under tensile-tensile loading mode. The testing of the implants is usually provided under the bending mode. Even though different testing condition for the conventional specimens tests and implants testing was adopted, a numerical study revealed their comparable fatigue properties. The fatigue limit determined for the implants was 105% higher than the one for coarse-grained and only by 4 % lower than the one for ultrafine-grained Ti grade 2. Based on the obtained results, conventional specimens testing can be used for the prediction of the fatigue limit of the implants.
    PracovištěÚstav fyziky materiálu
    KontaktYvonna Šrámková, sramkova@ipm.cz, Tel.: 532 290 485
    Rok sběru2022
    Elektronická adresahttps://www.sciencedirect.com/science/article/pii/S175161612100374X?via%3Dihub
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.