Počet záznamů: 1  

A Novel Statistical Model for Predicting the Efficacy of Vagal Nerve Stimulation in Patients With Epilepsy (Pre-X-Stim) Is Applicable to Different EEG Systems

  1. 1.
    SYSNO ASEP0543276
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevA Novel Statistical Model for Predicting the Efficacy of Vagal Nerve Stimulation in Patients With Epilepsy (Pre-X-Stim) Is Applicable to Different EEG Systems
    Tvůrce(i) Koriťáková, E. (CZ)
    Doležalová, I. (CZ)
    Chládek, Jan (UPT-D) RID, ORCID, SAI
    Jurková, T. (CZ)
    Chrastina, J. (CZ)
    Plešinger, Filip (UPT-D) RID, ORCID, SAI
    Roman, R. (CZ)
    Pail, M. (CZ)
    Jurák, Pavel (UPT-D) RID, ORCID, SAI
    Shaw, D. J. (CZ)
    Brázdil, M. (CZ)
    Celkový počet autorů11
    Číslo článku635787
    Zdroj.dok.Frontiers in Neuroscience
    Roč. 15, 11 May (2021)
    Poč.str.6 s.
    Forma vydáníOnline - E
    Jazyk dok.eng - angličtina
    Země vyd.CH - Švýcarsko
    Klíč. slovavagal nerve stimulation ; neurostimulation ; epilepsy ; efficacy prediction ; EEG reactivity ; epilepsy treatment
    Vědní obor RIVFH - Neurologie, neurochirurgie, neurovědy
    Obor OECDNeurosciences (including psychophysiology
    CEPNV19-04-00343 GA MZd - Ministerstvo zdravotnictví
    Způsob publikováníOpen access
    Institucionální podporaUPT-D - RVO:68081731
    UT WOS000653635600001
    EID SCOPUS85107209187
    DOI10.3389/fnins.2021.635787
    AnotaceBackground: Identifying patients with intractable epilepsy who would benefit from therapeutic chronic vagal nerve stimulation (VNS) preoperatively remains a major clinical challenge. We have developed a statistical model for predicting VNS efficacy using only routine preimplantation electroencephalogram (EEG) recorded with the TruScan EEG device (Brazdil et al., 2019). It remains to be seen, however, if this model can be applied in different clinical settings.
    Objective: To validate our model using EEG data acquired with a different recording system.
    Methods: We identified a validation cohort of eight patients implanted with VNS, whose preimplantation EEG was recorded on the BrainScope device and who underwent the EEG recording according to the protocol. The classifier developed in our earlier work, named Pre-X-Stim, was then employed to classify these patients as predicted responders or non-responders based on the dynamics in EEG power spectra. Predicted and real-world outcomes were compared to establish the applicability of this classifier. In total, two validation experiments were performed using two different validation approaches (single classifier or classifier voting).
    Results: The classifier achieved 75% accuracy, 67% sensitivity, and 100% specificity. Only two patients, both real-life responders, were classified incorrectly in both validation experiments.
    Conclusion: We have validated the Pre-X-Stim model on EEGs from a different recording system, which indicates its application under different technical conditions. Our approach, based on preoperative EEG, is easily applied and financially undemanding and presents great potential for real-world clinical use.
    PracovištěÚstav přístrojové techniky
    KontaktMartina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178
    Rok sběru2022
    Elektronická adresahttps://www.frontiersin.org/articles/10.3389/fnins.2021.635787/full