Počet záznamů: 1
Inconsistencies of neutron flux parameters for k(0) standardization in neutron activation analysis determined with the use of Au+Zr and Au+Mo+Cr monitor sets at the LVR-15 reactor in Rez
- 1.0384333 - ÚJF 2013 RIV HU eng J - Článek v odborném periodiku
Kubešová, Marie - Kučera, Jan
Inconsistencies of neutron flux parameters for k(0) standardization in neutron activation analysis determined with the use of Au+Zr and Au+Mo+Cr monitor sets at the LVR-15 reactor in Rez.
Journal of Radioanalytical and Nuclear Chemistry. Roč. 293, č. 2 (2012), s. 665-674. ISSN 0236-5731. E-ISSN 1588-2780
Grant CEP: GA ČR GA202/09/0363
Institucionální podpora: RVO:61389005
Klíčová slova: Neutron activation analysis * K(0) standardization * Neutron flux parameters
Kód oboru RIV: BG - Jaderná, atomová a mol. fyzika, urychlovače
Impakt faktor: 1.467, rok: 2012
Multipurpose research reactors require monitoring of neutron flux parameters (alpha, f, F (c) or I broken vertical bar (th) ) during every irradiation of samples when k(0) standardization in neutron activation analysis is to be used, because the above parameters may change quite frequently and unpredictably. We tested two monitor sets consisting of Au+Zr and Au+Mo+Cr for determination of the neutron flux parameters in k(0)-NAA at the LVR-15 reactor in A similar to eA3/4. Based on analysis of several reference materials, we found the use of the Au+Zr monitor set superior over the Au+Mo+Cr monitor set in the specific conditions of the LVR-15 reactor.
Trvalý link: http://hdl.handle.net/11104/0007250
Počet záznamů: 1