Počet záznamů: 1  

Time-Efficient Perfusion Imaging Using DCE- and DSC-MRI

  1. 1.
    SYSNO ASEP0497932
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevTime-Efficient Perfusion Imaging Using DCE- and DSC-MRI
    Tvůrce(i) Macíček, Ondřej (UPT-D) RID, ORCID, SAI
    Jiřík, Radovan (UPT-D) RID, ORCID, SAI
    Mikulka, J. (CZ)
    Bartoš, Michal (UTIA-B) ORCID
    Šprláková, A. (CZ)
    Keřkovský, M. (CZ)
    Starčuk jr., Zenon (UPT-D) RID, ORCID, SAI
    Bartušek, Karel (UPT-D) RID, ORCID, SAI
    Taxt, T. (NO)
    Celkový počet autorů9
    Zdroj.dok.Measurement Science Review. - : Sciendo - ISSN 1335-8871
    Roč. 18, č. 6 (2018), s. 262-271
    Poč.str.10 s.
    Forma vydáníTištěná - P
    Jazyk dok.eng - angličtina
    Země vyd.SK - Slovensko
    Klíč. slovaperfusion imaging ; contrast agents ; brain tumors ; DCE-MRI ; DSC-MRI
    Vědní obor RIVFS - Lékařská zařízení, přístroje a vybavení
    Obor OECDMedical engineering
    Vědní obor RIV – spolupráceÚstav teorie informace a automatizace - Využití počítačů, robotika a její aplikace
    CEPGA16-13830S GA ČR - Grantová agentura ČR
    LO1212 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    ED0017/01/01 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    LM2015062 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    EF16_013/0001775 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    Institucionální podporaUPT-D - RVO:68081731 ; UTIA-B - RVO:67985556
    UT WOS000452196100006
    EID SCOPUS85058653636
    DOI10.1515/msr-2018-0036
    AnotaceDynamic contrast enhanced MRI (DCE-MRI) and dynamic susceptibility contrast MRI (DSC-MRI) are perfusion imaging techniques used mainly for clinical and preclinical measurement of vessel permeability and capillary blood flow, respectively. It is advantageous to apply
    both methods to exploit their complementary information about the perfusion status of the tissue. We propose a novel acquisition method that combines advantages of the current simultaneous and sequential acquisition. The proposed method consists of a DCE-MRI acquisition
    interrupted by DSC-MRI acquisition. A new method for processing of the DCE-MRI data is proposed which takes the interleaved acquisition into account. Analysis of both the DCE- and DSC-MRI data is reformulated so that they are approximated by the same pharmacokinetic model (constrained distributed capillary adiabatic tissue homogeneity model). This provides a straightforward evaluation of the methodology as some of the estimated DCE- and DSC-MRI perfusion parameters should be identical. Evaluation on synthetic data showed an acceptable precision and no apparent bias introduced by the interleaved character of the DCE-MRI acquisition. Intravascular perfusion parameters obtained from clinical glioma data showed a fairly high correlation of blood flow estimates from DCE- and DSC-MRI, however, an unknown scaling factor was still present mainly because of the tissue-specific r*2 relaxivity. The results show validity of the proposed acquisition
    method. They also indicate that simultaneous processing of both DCE- and DSC-MRI data with joint estimation of some perfusion parameters (included in both DCE- and DSC-MRI) might be possible to increase the reliability of the DCE- and DSC-MRI methods alone.
    PracovištěÚstav přístrojové techniky
    KontaktMartina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178
    Rok sběru2019
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.