Time-Efficient Perfusion Imaging Using DCE- and DSC-MRI
- 1.
SYSNO ASEP 0497932 Druh ASEP J - Článek v odborném periodiku Zařazení RIV J - Článek v odborném periodiku Poddruh J Článek ve WOS Název Time-Efficient Perfusion Imaging Using DCE- and DSC-MRI Tvůrce(i) Macíček, Ondřej (UPT-D) RID, ORCID, SAI
Jiřík, Radovan (UPT-D) RID, ORCID, SAI
Mikulka, J. (CZ)
Bartoš, Michal (UTIA-B) ORCID
Šprláková, A. (CZ)
Keřkovský, M. (CZ)
Starčuk jr., Zenon (UPT-D) RID, ORCID, SAI
Bartušek, Karel (UPT-D) RID, ORCID, SAI
Taxt, T. (NO)Celkový počet autorů 9 Zdroj.dok. Measurement Science Review. - : Sciendo - ISSN 1335-8871
Roč. 18, č. 6 (2018), s. 262-271Poč.str. 10 s. Forma vydání Tištěná - P Jazyk dok. eng - angličtina Země vyd. SK - Slovensko Klíč. slova perfusion imaging ; contrast agents ; brain tumors ; DCE-MRI ; DSC-MRI Vědní obor RIV FS - Lékařská zařízení, přístroje a vybavení Obor OECD Medical engineering Vědní obor RIV – spolupráce Ústav teorie informace a automatizace - Využití počítačů, robotika a její aplikace CEP GA16-13830S GA ČR - Grantová agentura ČR LO1212 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy ED0017/01/01 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy LM2015062 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy EF16_013/0001775 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy Institucionální podpora UPT-D - RVO:68081731 ; UTIA-B - RVO:67985556 UT WOS 000452196100006 EID SCOPUS 85058653636 DOI 10.1515/msr-2018-0036 Anotace Dynamic contrast enhanced MRI (DCE-MRI) and dynamic susceptibility contrast MRI (DSC-MRI) are perfusion imaging techniques used mainly for clinical and preclinical measurement of vessel permeability and capillary blood flow, respectively. It is advantageous to apply
both methods to exploit their complementary information about the perfusion status of the tissue. We propose a novel acquisition method that combines advantages of the current simultaneous and sequential acquisition. The proposed method consists of a DCE-MRI acquisition
interrupted by DSC-MRI acquisition. A new method for processing of the DCE-MRI data is proposed which takes the interleaved acquisition into account. Analysis of both the DCE- and DSC-MRI data is reformulated so that they are approximated by the same pharmacokinetic model (constrained distributed capillary adiabatic tissue homogeneity model). This provides a straightforward evaluation of the methodology as some of the estimated DCE- and DSC-MRI perfusion parameters should be identical. Evaluation on synthetic data showed an acceptable precision and no apparent bias introduced by the interleaved character of the DCE-MRI acquisition. Intravascular perfusion parameters obtained from clinical glioma data showed a fairly high correlation of blood flow estimates from DCE- and DSC-MRI, however, an unknown scaling factor was still present mainly because of the tissue-specific r*2 relaxivity. The results show validity of the proposed acquisition
method. They also indicate that simultaneous processing of both DCE- and DSC-MRI data with joint estimation of some perfusion parameters (included in both DCE- and DSC-MRI) might be possible to increase the reliability of the DCE- and DSC-MRI methods alone.Pracoviště Ústav přístrojové techniky Kontakt Martina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178 Rok sběru 2019