Počet záznamů: 1
Density and sound velocity measurement by an Anton Paar DSA 5000 density meter: Precision and long-time stability.
- 1.0539386 - ÚCHP 2022 RIV NL eng J - Článek v odborném periodiku
Wagner, Zdeněk - Bendová, Magdalena - Rotrekl, Jan - Sýkorová, Adéla - Čanji, Maja - Parmar, Nirmal
Density and sound velocity measurement by an Anton Paar DSA 5000 density meter: Precision and long-time stability.
Journal of Molecular Liquids. Roč. 329, MAY 2021 (2021), č. článku 115547. ISSN 0167-7322. E-ISSN 1873-3166
Grant CEP: GA ČR(CZ) GA17-19798S
Institucionální podpora: RVO:67985858
Klíčová slova: density * viscosity correction * sound velocity
Obor OECD: Physical chemistry
Impakt faktor: 6.633, rok: 2021
Způsob publikování: Open access s časovým embargem
Vibrating tube density meters are frequently used in routine measurements. The instruments are often equiped with a cell for sound velocity measurement. The methodology of measurement is described in the operation manual and many papers contain recommedations how to achieve data with higher precision. This work is aimed to the study of importance of each aspect of the measurement procedures, effects of neglection of a viscosity correction as well as the magnitude of induced errors. Numerical values of all effects are presented. As a result of thorough analysis not yet published recommendations for best practice of density and sound velocity measurements are summarized.
Trvalý link: http://hdl.handle.net/11104/0317291
Počet záznamů: 1