Počet záznamů: 1  

First partial pressure measurements using Optical Penning Gauge in the COMPASS tokamak

  1. 1.
    SYSNO ASEP0617281
    Druh ASEPO - Ostatní výsledky
    Zařazení RIVO - Ostatní
    NázevFirst partial pressure measurements using Optical Penning Gauge in the COMPASS tokamak
    Tvůrce(i) Weinzettl, Vladimír (UFP-V) RID, ORCID
    Vartanian, S. (FR)
    Dejarnac, Renaud (UFP-V) RID, ORCID
    Jeřáb, Martin (UFP-V) ORCID
    Havránek, Aleš (UFP-V) ORCID
    Jaulmes, Fabien (UFP-V) ORCID
    Dimitrova, Miglena (UFP-V) ORCID
    Naydenkova, Diana (UFP-V) RID, ORCID
    Celkový počet autorů8
    Rok vydání2024
    Jazyk dok.eng - angličtina
    Země vyd.CZ - Česká republika
    Klíč. slovaCOMPASS ; measurement ; tokamak
    Vědní obor RIVBL - Fyzika plazmatu a výboje v plynech
    Obor OECDFluids and plasma physics (including surface physics)
    Institucionální podporaUFP-V - RVO:61389021
    AnotaceThe Optical Penning Gauge (OPG) for partial pressure measurements has been successfully implemented in the bottom high-field side diagnostic port in the COMPASS tokamak. The OPG diagnostic was based on an Alcatel CF2P type and two photomultipliers covered with interference filters as fast detectors. The procedure of its implementation on COMPASS is described in detail, mentioning technical improvements such as a new 3D printed light-proof shield and fine optics refocusing, leading to a doubling of the light collection efficiency, demonstrated on the measured calibration curves. Calibration data for deuterium (using 486 nm), helium (using 728 nm), nitrogen (around 820 nm) and argon (around 810 nm) are discussed, including the effect of deuterium pressure on the helium channel. Time-resolved measurements of the working gas (deuterium) and main gaseous impurity (helium) show, how helium, used for glow discharge cleaning, is subsequently released from the plasma-facing components during the following shot. Also, a typical behaviour of both the gasses in L-mode, ELM-free and ELMy H-mode regimes, and even just after the main (high-current high-temperature) discharge is exemplified. Surprisingly long, low-current post-discharges (2-3 kA, about 100 ms), predominantly in helium, were observed by the OPG but further also confirmed by an overview high-speed visible camera and by photomultipliers focused to the plasma edge. The OPG results were compared with a wide-range pressure gauge installed at the miplane on the low-field side, revealing similarities but also differences in both the time behaviour and magnitude. The values of the tokamak vacuum vessel pre-fill and the following gas puff, both extracted either from the calibrated gas puff piezo-valve or from the above-mentioned gauges, are compared to those required by the DYON code to achieve reasonable agreement with the measured plasma current, density and temperature for breakdown and burn-through discharge phases.
    PracovištěÚstav fyziky plazmatu
    KontaktVladimíra Kebza, kebza@ipp.cas.cz, Tel.: 266 052 975
    Rok sběru2025
    Elektronická adresahttps://soft2024.eu/wp-content/uploads/2024/09/SOFT-2024_Book-of-Abstracts_20-09-2024.pdf