Počet záznamů: 1  

Precise Measurement of T2 Using the Turbo FLASH Method

  1. 1.
    0399962 - ÚPT 2014 RIV SK eng C - Konferenční příspěvek (zahraniční konf.)
    Kořínek, Radim - Bartušek, Karel
    Precise Measurement of T2 Using the Turbo FLASH Method.
    Measurement 2013. Proceedings of the 9th International Conference on Measurement. Bratislava: Institute of Measurement Science SAS, 2013, s. 191-194. ISBN 978-80-969672-5-4.
    [Measurement 2013. International Conference on Measurement /9./. Smolenice (SK), 27.05.2013-30.05.2013]
    Grant CEP: GA ČR GAP102/11/0318; GA MŠMT ED0017/01/01
    Institucionální podpora: RVO:68081731
    Klíčová slova: Turbo FLASH * Relaxation T1 and T2 * Measurement * Relaxometry
    Kód oboru RIV: JA - Elektronika a optoelektronika, elektrotechnika

    The authors discuss the problem of using the Turbo FLASH sequence for the measurement of T2 in samples with short relaxation T1. In the case of the Turbo FLASH sequence, the relaxation T1 plays an essential role during the measurement of the T2 relaxation. The verification of the correct value of the T2 relaxation was performed using the spin echo (SE) approach with precise adjustment (RF flip angles). In order to clarify the errors occurring in the course of the T2 measurement, the authors used the ROMAG software (by Zenon Starcuk jr.) to carry out the simulations for the preparatory part of Turbo FLASH.
    Trvalý link: http://hdl.handle.net/11104/0227132

     
     
Počet záznamů: 1  