Počet záznamů: 1  

Construction and Application of Flow Enzymatic Biosensor Based of Silver Solid Amalgam Electrode for Determination of Sarcosine

  1. 1.
    0446079 - ÚFCH JH 2016 RIV DE eng J - Článek v odborném periodiku
    Josypčuk, Oksana - Barek, J. - Josypčuk, Bohdan
    Construction and Application of Flow Enzymatic Biosensor Based of Silver Solid Amalgam Electrode for Determination of Sarcosine.
    Electroanalysis. Roč. 27, č. 11 (2015), s. 2559-2566. ISSN 1040-0397. E-ISSN 1521-4109
    Grant CEP: GA ČR GBP206/12/G151; GA ČR GAP206/11/1638
    Institucionální podpora: RVO:61388955
    Klíčová slova: biosensors * sarcosine * silver solid amalgam electrode
    Kód oboru RIV: CG - Elektrochemie
    Impakt faktor: 2.471, rok: 2015

    In this paper, the flow amperometric enzymatic biosensor based on polished silver solid amalgam electrode for determination of sarcosine in model sample under flow injection analysis conditions is presented. The biosensor works on principle of electrochemical detection of oxygen decrease during enzymatic reaction which is directly proportional to the concentration of sarcosine in sample. The whole preparation process takes about 3 h. The RSD of repeatability of 10 consecutive measurements is 1.6% (csarcosine=1.0x10-4 moldm-3). Under optimal conditions the calibration dependence was linear in the range 7.5x10-6 – 5.0x10-4 moldm-3 and limit of detection was 2.0x10-6 moldm-3.
    Trvalý link: http://hdl.handle.net/11104/0248062

     
     