Počet záznamů: 1
Automatic marker-free estimation methods for the axis of rotation in sub-micron X-ray computed tomography
- 1.0580354 - FZÚ 2024 RIV NL eng J - Článek v odborném periodiku
Zemek, M. - Šalplachta, J. - Zikmund, T. - Omote, K. - Takeda, Y. - Oberta, Peter - Kaiser, J.
Automatic marker-free estimation methods for the axis of rotation in sub-micron X-ray computed tomography.
Tomography of Materials and Structures. Roč. 1, Mar. (2023), č. článku 100002. ISSN 2949-673X
Grant CEP: GA MŠMT LM2018110
Institucionální podpora: RVO:68378271
Klíčová slova: computed tomography * rotation axis * tuning-fork artifact * automatic
Obor OECD: Optics (including laser optics and quantum optics)
Způsob publikování: Open access
Misalignment of the rotation axis causes severe artifacts in X-ray computed tomography. Calibration of this parameter is often insufficient for sub-micron resolution measurements and needs to be corrected during the post-processing. This correction can be accelerated by various automatic methods. These vary in mechanisms and performance, making them suitable for different use-cases. This work summarizes existing automatic methods for estimating the rotation axis in X-ray computed tomography, with a focus on sub-micron applications. Some of the methods are implemented and compared in the context of a laboratory sub-micron scanner to demonstrate practical considerations of this task.
Trvalý link: https://hdl.handle.net/11104/0349138
Název souboru Staženo Velikost Komentář Verze Přístup 0580354.pdf 1 7.5 MB CC licence Vydavatelský postprint povolen