Počet záznamů: 1
Measurement of the Hydrogen Cyanide Absorption Lines' Centers with the Potential for Mise en Pratique
- 1.0552425 - ÚPT 2022 RIV US eng C - Konferenční příspěvek (zahraniční konf.)
Hošek, Martin - Řeřucha, Šimon - Hrabina, Jan - Čížek, Martin - Číp, Ondřej
Measurement of the Hydrogen Cyanide Absorption Lines' Centers with the Potential for Mise en Pratique.
2021 Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS). Proceedings. New York: IEEE, 2021, (2021), č. článku 7189. ISBN 978-1-6654-3935-0.
[2021 Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS). online (US), 07.07.2021-17.07.2021]
Grant CEP: GA ČR(CZ) GA19-14988S; GA MŠMT(CZ) EF16_026/0008460
GRANT EU: European Commission(XE) 17IND03 - LaVa
Institucionální podpora: RVO:68081731
Klíčová slova: hydrogen cyanide * linear absorption spectroscopy * optical frequency comb * frequency reference * frequency stabilized lasers * SI meter
Obor OECD: Optics (including laser optics and quantum optics)
https://ieeexplore.ieee.org/document/9604261
We built a linear absorption spectroscopy setup to measure hydrogen cyanide (H13C14N) 2?3 rotational-vibrational band. The laser's frequency was locked to the minimum of the absorption line by using the 1f technique. We used a commercially available lock-in amplifier combined with a custom-made PID controller allowing for the short-term frequency stability of about 10-10. The laser frequency was determined by measuring the beat-note between the laser and the optical frequency comb. The laser was locked to each absorption line for about 8 hours and the final time record allowed us to determine the absorption line central frequency with the uncertainty of about 40 kHz. This shows significant improvement compared to the data already available for HCN. Our work can potentially lead to the acceptance of HCN for Mise en Pratique (MeP), which means that laser locked to one of the HCN lines can become the official source of the one-meter unit.
Trvalý link: http://hdl.handle.net/11104/0327556
Počet záznamů: 1