Physical properties of modern reciprocal endodontic systems and fatigue failure testing in simulated clinical conditions

  1. 1.
    0551283 - FZÚ 2022 RIV CH eng J - Článek v odborném periodiku
    Jusku, A. - Tomáštík, Jan - Václavek, Lukáš - Jirásek, P. - Harvan, L. - Holík, P. - Čtvrtlík, Radim
    Physical properties of modern reciprocal endodontic systems and fatigue failure testing in simulated clinical conditions.
    Applied Sciences-Basel. Roč. 11, č. 23 (2021), č. článku 11160. E-ISSN 2076-3417
    Grant CEP: GA TA ČR TN01000038
    Institucionální podpora: RVO:68378271
    Klíčová slova: cyclic fatigue * endodontic instruments * nanoindentation * roughness
    Obor OECD: Dentistry, oral surgery and medicine
    Impakt faktor: 2.838, rok: 2021
    Způsob publikování: Open access

    Instrument fracture ranks among the most crucial complications during the endodontic treatment of a tooth. In order to better understand the practical limits of the instrument, the relation between the cyclic fatigue resistance and physical properties such as hardness, modulus of elasticity, creep and surface roughness were explored. Cyclic fatigue testing in an artificial root canal at intracanal temperature, nanoindentation and 3D microscopy were used for evaluation of four commonly used thermomechanically treated NiTi endodontic instruments (Unicone Plus 6/025, Unicone 6/025, Reciproc Blue R25 and WaveOne Gold Primary). Cyclic fatigue results were analyzed using the Kruskal–Wallis, Mann–Whitney and Bonferroni corrections. The wear resistance of Unicone 6/025 instruments was significantly lower compared to all other instruments (p [removed] 0.05).

    Trvalý link: http://hdl.handle.net/11104/0326713

     
    Název souboruStaženoVelikostKomentářVerzePřístup
    0551283.pdf143.2 MBCC LicenceVydavatelský postprintpovolen
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.