Počet záznamů: 1  

Design of modular multi-channel electron spectrometers for application in laser matter interaction experiments at Prague Asterix Laser System

  1. 1.
    0548513 - ÚFP 2022 RIV US eng J - Článek v odborném periodiku
    Krupka, Michal - Singh, Sushil Kumar - Pisarczyk, T. - Dostál, Jan - Kálal, Milan - Krása, J. - Dudžák, Roman - Burian, Tomáš - Jelínek, Šimon - Chodukowski, T. - Rusiniak, Z. - Krůs, Miroslav - Juha, Libor
    Design of modular multi-channel electron spectrometers for application in laser matter interaction experiments at Prague Asterix Laser System.
    Review of Scientific Instruments. Roč. 92, č. 2 (2021), č. článku 023514. ISSN 0034-6748. E-ISSN 1089-7623
    Grant CEP: GA MŠMT(CZ) LM2015083; GA MŠMT(CZ) LM2018114; GA MŠMT EF16_013/0001552; GA ČR GA19-02545S; GA ČR(CZ) GA19-24619S
    GRANT EU: European Commission(XE) 654148 - LASERLAB-EUROPE; European Commission(XE) 633053 - EUROfusion
    Institucionální podpora: RVO:61389021
    Klíčová slova: PALS * electron spectrometer
    Obor OECD: Optics (including laser optics and quantum optics)
    Impakt faktor: 1.843, rok: 2021
    Způsob publikování: Omezený přístup
    https://aip.scitation.org/doi/10.1063/5.0029849

    This paper describes design, development, and implementation of a multi-channel magnetic electron spectrometer for the application in laser-plasma interaction experiments carried out at the Prague Asterix Laser System. Modular design of the spectrometer allows the setup in variable configurations to evaluate the angular distribution of hot electron emission. The angular array configuration of the electron spectrometers consists of 16 channels mounted around the target. The modules incorporate a plastic electron collimator designed to suppress the secondary radiation by absorbing the wide angle scattered electrons and photons inside the collimator. The compact model of the spectrometer measures electron energies in the range from 50 keV to 1.5MeV using ferrite magnets and from 250 keV to 5MeV using stronger neodymium magnets. An extended model of the spectrometer increases the measured energy range up to 21MeV or 35MeV using ferrite or neodymium magnets, respectively. Position to energy calibration was obtained using the particle tracking simulations. The experimental results show the measured angularly resolved electron energy distribution functions from interaction with solid targets. The angular distribution of hot electron temperature, the total flux, and the maximum electron energy show a directional dependence. The measured values of these quantities increase toward the target normal. For a copper target, the average amount of measured electron flux is 1.36 × 1011, which corresponds to the total charge of about 21 nC.
    Trvalý link: http://hdl.handle.net/11104/0324533

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.