Data acquisition system for the CALICE AHCAL calorimeter
- 1.0520471 - FZÚ 2020 RIV GB eng J - Článek v odborném periodiku
Kvasnička, Jiří
Data acquisition system for the CALICE AHCAL calorimeter.
Journal of Instrumentation. Roč. 12, č. 3 (2017), s. 1-10, č. článku C03043. ISSN 1748-0221. E-ISSN 1748-0221
Grant CEP: GA MŠMT LG14033
GRANT EU: European Commission(XE) 654168 - AIDA-2020
Institucionální podpora: RVO:68378271
Klíčová slova: CALICE * data acquisition * FPGA * calorimeter * performance * electronics: readout * electronics: design * calorimeter: hadronic
Obor OECD: Particles and field physics
Impakt faktor: 1.258, rok: 2017
Způsob publikování: Omezený přístup
https://doi.org/10.1088/1748-0221/12/03/c03043
The data acquisition system (DAQ) for a highly granular analogue hadron calorimeter (AHCAL) for the future International Linear Collider is presented. The developed DAQ chain has several stages of aggregation and scales up to 8 million channels foreseen for the AHCAL detector design. The largest aggregation device, Link Data Aggregator, has 96 HDMI connectors, four Kintex7 FPGAs and a central Zynq System-On-Chip. Architecture and performance results are shown in detail. Experience from DESY testbeams with a small detector prototype consisting of 15 detector layers are shown.
Trvalý link: http://hdl.handle.net/11104/0305142