Počet záznamů: 1  

Generating standardized image data for testing and calibrating quantification of volumes, surfaces, lengths, and object counts in fibrous and porous materials using X-ray microtomography

  1. 1.
    0497407 - ÚSMH 2019 RIV US eng J - Článek v odborném periodiku
    Jiřík, M. - Bartoš, M. - Tomášek, P. - Malečková, A. - Kural, T. - Horáková, J. - Lukáš, D. - Suchý, Tomáš - Kochová, P. - Hubálek Kalbáčová, M. - Králíčková, M. - Tonar, Z.
    Generating standardized image data for testing and calibrating quantification of volumes, surfaces, lengths, and object counts in fibrous and porous materials using X-ray microtomography.
    Microscopy Research and Technique. Roč. 81, č. 6 (2018), s. 551-568. ISSN 1059-910X. E-ISSN 1097-0029
    Grant CEP: GA MZd(CZ) NV15-25813A
    Institucionální podpora: RVO:67985891
    Klíčová slova: fibers * Python * pores * scaffolds * spatial statistics * stereology * textile
    Obor OECD: Computer hardware and architecture
    Impakt faktor: 1.327, rok: 2018
    https://onlinelibrary.wiley.com/doi/abs/10.1002/jemt.23011

    Quantification of the structure and composition of biomaterials using micro-CT requires image segmentation due to the low contrast and overlapping radioopacity of biological materials. The amount of bias introduced by segmentation procedures is generally unknown. We aim to develop software that generates three-dimensional models of fibrous and porous structures with known volumes, surfaces, lengths, and object counts in fibrous materials and to provide a software tool that calibrates quantitative micro-CT assessments. Virtual image stacks were generated using the newly developed software TeIGen, enabling the simulation of micro-CT scans of unconnected tubes, connected tubes, and porosities. A realistic noise generator was incorporated. Forty image stacks were evaluated using micro-CT, and the error between the true known and estimated data was quantified. Starting with geometric primitives, the error of the numerical estimation of surfaces and volumes was eliminated, thereby enabling the quantification of volumes and surfaces of colliding objects. Analysis of the sensitivity of the thresholding upon parameters of generated testing image sets revealed the effects of decreasing resolution and increasing noise on the accuracy of the micro-CT quantification. The size of the error increased with decreasing resolution when the voxel size exceeded 1/10 of the typical object size, which simulated the effect of the smallest details that could still be reliably quantified. Open-source software for calibrating quantitative micro-CT assessments by producing and saving virtually generated image data sets with known morphometric data was made freely available to researchers involved in morphometry of three-dimensional fibrillar and porous structures in micro-CT scans.
    Trvalý link: http://hdl.handle.net/11104/0289978

     
     
Počet záznamů: 1  