    Feasibility of in-situ trapping of selenium hydride in a DBD atomizer for ultrasensitive Se determination by atomic absorption spectrometry studied with 75Se radioactive indicator.
    Journal of Analytical Atomic Spectrometry. Roč. 34, č. 1 (2019), s. 193-202. ISSN 0267-9477. E-ISSN 1364-5544
    Grant CEP: GA ČR GA18-01116S
    Institucionální podpora: RVO:68081715
    Klíčová slova: hydride preconcentration * radiotracer * dielectric barrier discharge
    Obor OECD: Analytical chemistry
    Impakt faktor: 3.498, rok: 2019

    Preconcentration of selenium hydride in a planar dielectric barrier discharge (DBD) atomizer was optimized with detection by atomic absorption spectrometry. The use of 75Se radiotracer combined with radiometry and autoradiography allowed proving efficient analyte trapping (90%) in a small spot in the central part of the DBD optical arm whereas incomplete analyte release was found responsible for preconcentration efficiency of 70%.