Minimization of Theoretical Error of Input Parameters for Vapor Permeation Apparatus.
- 1.0493317 - ÚCHP 2019 RIV DE eng J - Článek v odborném periodiku
Dytrych, Pavel - Vajglová, Zuzana - Morávková, Lenka - Jandová, Věra - Izák, Pavel - Petrusová, Zuzana
Minimization of Theoretical Error of Input Parameters for Vapor Permeation Apparatus.
Chemical Engineering & Technology. Roč. 41, č. 9 (2018), s. 1727-1736. ISSN 0930-7516. E-ISSN 1521-4125
Grant CEP: GA ČR GJ17-03367Y
Institucionální podpora: RVO:67985858
Klíčová slova: error identification * membrane flux * hexane vapour permation
Obor OECD: Chemical process engineering
Impakt faktor: 2.418, rok: 2018
All input experimental errors for membrane flux determination were analyzed. The overall error in pressure did not largely influence the determination of membrane flux. The main contribution to errors was ascribed to flow meters and membrane thickness. The choice of an optimal operational range for input parameters led to elimination of at least 2/3 of the theoretical relative error for subsequent evaluation of the permeability of individual species. This is especially valuable for low-permeable membranes, whereas a very low relative theoretical error can be obtained in the case of highly permeable materials.
Trvalý link: http://hdl.handle.net/11104/0286695