Počet záznamů: 1  

Impact of the shape of the differentially pumped chamber on critical flow character

  1. 1.
    0490115 - ÚPT 2019 RIV US eng C - Konferenční příspěvek (zahraniční konf.)
    Hlavatá, P. - Maxa, J. - Bílek, M. - Neděla, Vilém
    Impact of the shape of the differentially pumped chamber on critical flow character.
    ECS Transactions. Pennington: Electrochemical Society, 2017, Roč. 81, č. 1 (2017), s. 317-322. ISBN 9781607685395. ISSN 1938-6737.
    [International Conference on Advanced Batteries, Accumulators and Fuel Cells, ABAF 2017 /18./. Brno (CZ), 10.09.2017-13.09.2017]
    Institucionální podpora: RVO:68081731
    Klíčová slova: critical flow * primary electron beams
    Obor OECD: Thermodynamics

    The Electrochemical Society. The article deals with the impact of the shape of the differential pumped chamber of the Environmental electron microscope on resultant value of average Normalized density number on the path of the primary electron beam which has directly proportional influence on the dispersion of this beam and thereby influence on sharpness resulting image. Analysis were made using finite volume method in Ansys Fluent system with the respect to critical flow created behind the aperture which separates differential pumped chamber from the specimen chamber.
    Trvalý link: http://hdl.handle.net/11104/0284444

     
     