Počet záznamů: 1  

Langerhans Islet Volume Estimation from 3D Optical Projection Tomography

  1. 1.
    0485780 - FGÚ 2018 RIV DE eng C - Konferenční příspěvek (zahraniční konf.)
    Švihlík, J. - Kybic, J. - Habart, D. - Hlushak, H. - Dvořák, J. - Radochová, Barbora
    Langerhans Islet Volume Estimation from 3D Optical Projection Tomography.
    Computer Vision – ACCV 2016 Workshops. Cham: Springer, 2017 - (Ma, K.; Lu, J.; Chen, C.), s. 583-594. Lecture Notes in Computer Science, 10117. ISBN 978-3-319-54426-7.
    [Asian Conference on Computer Vision /13./. Taipei (TW), 20.11.2016-24.11.2016]
    Grant CEP: GA MŠMT(CZ) LM2015062
    Institucionální podpora: RVO:67985823
    Klíčová slova: Langerhans islets * digital image analysis * optical projection tomography
    Obor OECD: Medical laboratory technology (including laboratory samples analysis

    This paper concerns the comparison of automatic volume estimation methods for isolated pancreatic islets. The estimated islet volumes are needed during the process of assessing the islet sample quality prior to the islet transplantation. We study several different methods for automatic volume estimation. For this purpose we acquired a set of projections using optical tomography for a sample of an islet population. Based on these projections we estimated the islet volumes using two stereological methods (the automatic Wulfsohn’s method and the manual fakir method, considered to be the ground truth in this study), together with the filtered back projection followed by 3D segmentation. We have also employed two simple methods, currently used in medical practice, based on fitting a sphere or a prolate ellipsoid to a single binarized 2D islet projections.
    Trvalý link: http://hdl.handle.net/11104/0280724

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.