Počet záznamů: 1  

Measurement of camera spectral response

  1. 1.
    0473020 - ÚFP 2017 RIV CZ eng V - Výzkumná zpráva
    Melich, Radek - Václavík, Jan - Žídek, Karel
    Measurement of camera spectral response.
    Turnov: ELCOM, a.s., 2016. 3 s.
    Zdroj financování: N - neveřejné zdroje
    Klíčová slova: Camera spectral response
    Kód oboru RIV: JA - Elektronika a optoelektronika, elektrotechnika

    Spectral response of the camera will be measured using monochromator in conjunction with an integration sphere. Camera without objective will be placed in one port of the integrating sphere and light beam from monochromator will be fed into perpendicular port. Sensor of the camera sensor will be therefore uniformly illuminated by diffused light from monochromator. Proposed FWHM of the monochromator is 15 nm as a trade-off between light intensity provided by monochromator and deformation of spectral characteristic (Picture XY). Diameter of the integrating sphere has to be larger than 100 mm to ensure uniform illumination over sensor area.