Počet záznamů: 1  

Development of antibacterial quaternary ammonium silane coatings on polyurethane catheters

  1. 1.
    0456882 - FZÚ 2016 RIV US eng J - Článek v odborném periodiku
    Zanini, S. - Polissi, A. - Maccagni, E.A. - Dell'Orto, E.C. - Liberatore, Chiara - Riccardi, C.
    Development of antibacterial quaternary ammonium silane coatings on polyurethane catheters.
    Journal of Colloid and Interface Science. Roč. 451, Aug (2015), 78-84. ISSN 0021-9797. E-ISSN 1095-7103
    Grant CEP: GA MŠMT EE2.3.20.0143
    Grant ostatní: OP VK 6(XE) CZ.1.07/2.3.00/20.0143
    Institucionální podpora: RVO:68378271
    Klíčová slova: plasma-induced graft-polymerization * acrylic acid * ATR/FTIR * AFM * quaternary ammonium silane * Escherichia coli
    Kód oboru RIV: BH - Optika, masery a lasery
    Impakt faktor: 3.782, rok: 2015

    Polyurethene (PU) catheters were coated with 3-(trimethoxysilyl)-propyldimethyloctadecylammonium chloride (QAS) by means of a multistep process which involved a vapor phase plasma-induced graft-polymerization of acrylic acid (AAc). The AAc coating, whose stability in aqueous media was assessed by immersion in Phospate Buffer Saline (PBS), was characterized by means of Attenuated Total Reflectance Fourier Transform Infrared (ATR/FTIR) spectroscopy. Moreover, the COOH surface density was evaluated by a colorimetric assay with Methylene Blue. Carrying a negative charge at neutral pH, AAc coatings were proficient in positively charged molecules (like QAS) adsorption. ATR/FTIR spectroscopy and a colorimetric assay with Bromophenol Blue allowed us to verify the presence and the uniformity of the QAS coating on the PU catheters and the positive effect of the AAc graft-polymerization on the QAS adsorption.

    Trvalý link: http://hdl.handle.net/11104/0257356

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.