Počet záznamů: 1  

Thomson spectrometer–microchannel plate assembly calibration for MeV-range positive and negative ions, and neutral atoms

  1. 1.
    0439126 - FZÚ 2015 RIV US eng J - Článek v odborném periodiku
    Prasad, R. - Abicht, F. - Borghesi, Marco - Braenzel, J. - Nickles, P.V. - Priebe, G. - Schnürer, M. - Ter-Avetisyan, Sargis
    Thomson spectrometer–microchannel plate assembly calibration for MeV-range positive and negative ions, and neutral atoms.
    Review of Scientific Instruments. Roč. 84, č. 5 (2013), "053302-1"-"053302-6". ISSN 0034-6748. E-ISSN 1089-7623
    Grant CEP: GA MŠMT ED1.1.00/02.0061; GA MŠMT EE2.3.20.0279
    Grant ostatní: ELI Beamlines(XE) CZ.1.05/1.1.00/02.0061; LaserZdroj (OP VK 3)(XE) CZ.1.07/2.3.00/20.0279
    Institucionální podpora: RVO:68378271
    Klíčová slova: implantation * beam
    Kód oboru RIV: BL - Fyzika plazmatu a výboje v plynech
    Impakt faktor: 1.584, rok: 2013

    We report on the absolute calibration of a microchannel plate (MCP) detector, used in conjunction with a Thomson parabola spectrometer. The calibration delivers the relation between a registered count numbers in the CCD camera (on which the MCP phosphor screen is imaged) and the number of ions incident on MCP. The particle response of the MCP is evaluated for positive, negative, and neutral particles at energies below 1 MeV.
    Trvalý link: http://hdl.handle.net/11104/0242435

     
     
Počet záznamů: 1  