Počet záznamů: 1  

A miniaturized device for bioluminescence analysis of caspase-3/7 activity in a single apoptotic cell

  1. 1.
    0437556 - ÚŽFG 2015 RIV DE eng J - Článek v odborném periodiku
    Adamová, Eva - Lišková, M. - Matalová, Eva - Klepárník, K.
    A miniaturized device for bioluminescence analysis of caspase-3/7 activity in a single apoptotic cell.
    Analytical and Bioanalytical Chemistry. Roč. 406, č. 22 (2014), s. 5389-5394. ISSN 1618-2642. E-ISSN 1618-2650
    Grant CEP: GA ČR GAP304/11/1418
    Institucionální podpora: RVO:67985904
    Klíčová slova: apoptosis * bioluminiscence * caspase3/7 * single-cell analysis
    Kód oboru RIV: CE - Biochemie
    Impakt faktor: 3.436, rok: 2014

    Caspases are key enzymes activated during the apoptotic machinery. Apoptosis as a way of programmed cell death becomes deregulated in some pathologies including cancer transformations, neurodegenerative, or autoimmune diseases. Most of the methods available for the detection of apoptosis and caspases provide qualitative information only or quantification data as an average from cell populations or cell lysates. Several reports point to the importance of more accurate single-cell analyses in biomedical studies due to heterogeneity at tissue as well as cell level. To meet these requirements, we developed a miniaturized device enabling detection and quantification of active caspase-3/7 in individual cells at a femtogram level (10(-15) g). The active caspase-3/7 detection protocol is based on the bioluminescence chemistry commercially available as a Caspase-Glo (TM) 3/7 reagent developed by Promega. As a model, we used human stem cells treated by camptothecin to induce apoptosis. Individual apoptotic cells were captured from a culture medium under a microscope and transferred by a micromanipulation system into a detection capillary containing 2 mu l of the reagent. Cells without activation by camptothecin served as negative controls. The detection limit of active caspase-3/7 achieved in the miniaturized system was determined as 0.20 and limit of quantification as 0.65 of the amount found in a single apoptotic human stem cell. Such a sensitive method could have a wide application potential in laboratory medicine and related clinically oriented research.
    Trvalý link: http://hdl.handle.net/11104/0241099

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.