Počet záznamů: 1
Silver behenate and silver stearate powders for calibration of SAS instruments
- 1.0391898 - ÚMCH 2014 RIV GB eng J - Článek v odborném periodiku
Nyam-Osor, M. - Soloviov, D. V. - Kovalev, Yu. S. - Zhigunov, Alexander - Rogachev, A. V. - Ivankov, O. I. - Erhan, R. V. - Kuklin, A. I.
Silver behenate and silver stearate powders for calibration of SAS instruments.
Journal of Physics: Conference Series. Roč. 351, č. 1 (2012), 012024_1-012024_9. ISSN 1742-6588. E-ISSN 1742-6596.
[SANS-YuMO User Meeting. Dubna, 27.05.2011-30.05.2011]
Výzkumný záměr: CEZ:AV0Z40500505
Klíčová slova: silver behenate * silver stearate * SAS
Kód oboru RIV: CD - Makromolekulární chemie
The possibilities of calibrations by silver behenate [CH3(CH2)(20)COOAg] and silver stearate [CH3 (CH2)(16)CO2Ag] for the small angle scattering (SAS) technique are considered. It was shown that the long-period spacing of the silver compounds allow us to check the parameters of time-of-flight (TOF) method. Results of the data obtained from small angle neutron scattering (SANS) measurements performed with the silver behenate and the silver stearate powders are presented and compared with small angle X-ray scattering (SAXS) measurements.
Trvalý link: http://hdl.handle.net/11104/0222038