Počet záznamů: 1  

Rotary slanted single wire CTA - a useful tool for 3D flows investigations

  1. 1.
    0386098 - ÚT 2013 RIV CZ eng C - Konferenční příspěvek (zahraniční konf.)
    Jonáš, Pavel
    Rotary slanted single wire CTA - a useful tool for 3D flows investigations.
    Experimental Fluid Mechanics 2012. Liberec: Technical University of Liberec, 2012 - (Vít, T.; Dančová, P.; Novotný, P.), s. 318-323. ISBN 978-80-7372-912-7.
    [Experimental Fluid Mechanics 2012 /7./. Hradec Králové (CZ), 20.11.2012-23.11.2012]
    Grant CEP: GA ČR GAP101/12/1271; GA ČR GA101/08/1112
    Institucionální podpora: RVO:61388998
    Klíčová slova: hot wire probe * constant temperature anemometer * 3D flow
    Kód oboru RIV: BK - Mechanika tekutin

    The procedure is described of experimental investigation of a statistically stationary generally nonisothermal 3D flow by means of a constant temperature anemometer (CTA) using single slanted heated wire, rotary round the fixed axis. The principle of this procedure is quite clear. The change of the heated wire temperature modifies ratio of CTA sensitivities to temperature and velocity fluctuations. Turning the heated wire through a proper angle changes the sensitivity to components of the instantaneous velocity vector. Some recommendations are presented based on long time experiences, e.g. on the choice of probe, on the probe calibration, to the measurement organization and to the evaluation of results.