Počet záznamů: 1
Analytical Value of Detecting an Individual Molecular Binding Event:The Case of the Surface Plasmon Resonance Biosensor
- 1.0374553 - ÚFE 2012 RIV US eng J - Článek v odborném periodiku
Šípová, Hana - Homola, Jiří
Analytical Value of Detecting an Individual Molecular Binding Event:The Case of the Surface Plasmon Resonance Biosensor.
Analytical Chemistry. Roč. 84, č. 1 (2012), s. 30-34. ISSN 0003-2700. E-ISSN 1520-6882
Grant CEP: GA AV ČR KAN200670701; GA MŠMT OC09058; GA MŠMT(CZ) LH11102
Výzkumný záměr: CEZ:AV0Z20670512
Klíčová slova: surface plasmon resonance * single-molecule detection * kinetics * microfluidics
Kód oboru RIV: JB - Senzory, čidla, měření a regulace
Impakt faktor: 5.695, rok: 2012
In this study, we compare the bioanalytical performance of flow-through surface plasmon resonance (SPR) sensors based on (1) localized surface plasmons (LSP) which detect a single binding event and (2) propagating surface plasmons (PSP) which integrate a great number of simultaneously occurring binding events. We demonstrate that for the biomolecular interactions most relevant to biosensing the LSP approach is inferior to the PSP approach due to low probability of positive sensor response.
Trvalý link: http://hdl.handle.net/11104/0207450
Název souboru Staženo Velikost Komentář Verze Přístup UFE 0374553.pdf 6 590.9 KB Jiná vyžádat
Počet záznamů: 1