Počet záznamů: 1
Distribution of polyethylene wear particles and bone fragments in periprosthetic tissue around total hip joint replacements
- 1.0347024 - ÚMCH 2011 RIV GB eng J - Článek v odborném periodiku
Zolotarevova, E. - Entlicher, G. - Pavlova, Ewa - Šlouf, Miroslav - Pokorný, D. - Veselý, F. - Gallo, J. - Sosna, A.
Distribution of polyethylene wear particles and bone fragments in periprosthetic tissue around total hip joint replacements.
Acta Biomaterialia. Roč. 6, č. 9 (2010), s. 3595-3600. ISSN 1742-7061. E-ISSN 1878-7568
Grant CEP: GA MŠMT 2B06096
Výzkumný záměr: CEZ:AV0Z40500505
Klíčová slova: joint replacement * polyethylene * wear particles distribution
Kód oboru RIV: CD - Makromolekulární chemie
Impakt faktor: 4.824, rok: 2010
Ultra-high molecular weight polyethylene (UHMWPE) wear particles play a significant role in failures of total joint replacements (TJRs). In this work, we investigated the distribution of these wear particles in periprosthetic tissues obtained from nine revisions of hip TJR. All periprosthetic tissues were combined and mechanically separated into granuloma tissue (containing hard granules visible to the naked eye) and surrounding tissue (without visible granules). Of the total amount of polyethylene particles in the whole granuloma tissue, 72% of particles in the size range 0.1–10 um and 68% of those larger than 10 um were found in granules. Therefore, the formation of granules significantly lowers the effective amount of wear particles available for interaction with reactive cells and seems to be a natural defense mechanism.
Trvalý link: http://hdl.handle.net/11104/0006002