Počet záznamů: 1  

Stress and reliability analyses of ceramic femoral heads with axisymmetric production inaccuracies

  1. 1.
    0211902 - UT-L 20010154 RIV HR eng C - Konferenční příspěvek (zahraniční konf.)
    Fuis, Vladimír - Janíček, P.
    Stress and reliability analyses of ceramic femoral heads with axisymmetric production inaccuracies.
    Mediterranean Conference on Medical and Biological Engineering and Computing. Pula: Faculty of Electrical Engineering, Zagreb, 2001 - (Magjarevič, R.; Tonkovič, S.; Bilas, V.; Lackovič, I.), s. 632-635. Part. ISBN 953-184-024-5. ISSN 1680-0737.
    [IX Mediteranean Conference on Medical and Biological Engineering and Computing. Pula (HR), 12.06.2001-15.06.2001]
    Grant CEP: GA ČR GA101/01/0974
    Klíčová slova: hip joint endoprosthesis * production inaccuracy * ceramic head
    Kód oboru RIV: JH - Keramika, žáruvzdorné materiály a skla

    The paper contains the results of the stress and reliability analyses of the ceramic head of the total hip joint endoprosthesis under testing conditions in accordance with the norm ISO 7206-5. The axisymmetric production inaccuracies of the contact ta** of the head and the stem are considered. The effects of the production inaccuracies on the head's hoop stress and the head's contact pressure along the contact taper are analyzed in detail.
    Trvalý link: http://hdl.handle.net/11104/0107451

     
     