Optimization of nickel tetracarbonyl generation and nickel in situ trapping within a transversely heated graphite furnace for atomic absorption spectrometry

  1. 1.
    0162186 - UIACH-O 200021 RIV NL eng J - Článek v odborném periodiku
    Dočekal, Bohumil - Marek, Petr
    Optimization of nickel tetracarbonyl generation and nickel in situ trapping within a transversely heated graphite furnace for atomic absorption spectrometry.
    Spectrochimica Acta Part B: Atomic Spectroscopy. Roč. 55, - (2000), s. 535-543. ISSN 0584-8547. E-ISSN 1873-3565
    Grant CEP: GA ČR GA203/98/0754
    Kód oboru RIV: CB - Analytická chemie, separace
    Impakt faktor: 2.608, rok: 2000

    Influence of experimental parameters of flow injection generation and in situ collection within transversely heated graphite furnace for atomic absorption spectrometric determination of nickel was investigated. Acidity of the reaction mixture, kinetics of carbonyl formation, the temperature of the tube, gas flow rate and the position of the capillary are the most critical parameters. Under optimum conditions, an overall efficiency of about 80 percent can be achieved.