Počet záznamů: 1  

Comparison of fluorine determination in biological and environmental samples by NAA, PAA and PIXE

  1. 1.
    0101848 - UJF-V 20043015 RIV NL eng J - Článek v odborném periodiku
    Havránek, Vladimír - Kučera, Jan - Řanda, Zdeněk - Voseček, Václav
    Comparison of fluorine determination in biological and environmental samples by NAA, PAA and PIXE.
    [Porovnání určení fluoru v biologických a ekologických vzorcích.]
    Journal of Radioanalytical and Nuclear Chemistry. Roč. 259, č. 2 (2004), s. 275-282. ISSN 0236-5731. E-ISSN 1588-2780
    Grant CEP: GA ČR GV202/97/K038; GA ČR GA104/00/1297
    Výzkumný záměr: CEZ:AV0Z1048901
    Klíčová slova: activation-analysis
    Kód oboru RIV: DN - Vliv životního prostředí na zdraví
    Impakt faktor: 0.457, rok: 2004

    Capabilities of instrumental neutron activation analysis (INAA), photon activation analysis with radiochemical separation (RPAA) and proton induced gamma-ray emission (PIGE) for determination of fluorine in biological and environmental materials are compared in terms of detection limits, nuclear and other interferences. The lowest detection limit of 0.5 mug.g(-1) was achieved by RPAA, the detection limit of the non destructive and interference-free PIGE method was by one order of magnitude higher, whereas two to three orders of magnitude higher detection limits compared to RPAA were obtained in INAA. A satisfactory agreement was found between the results obtained in this work and reference, mostly information values for several biological and environmental materials. However, higher values of two orders of magnitude were determined by RPAA and PIGE in NIST RM-8414 Bovine Muscle compared to the fluorine information value. The reason for this difference is the contamination of this material with Teflon, which occurred on its preparation and has not been taken into account in establishing the information value.

    Možnosti instrumentální neutronové aktivační analýzy (INAA), fotonové aktivační analýzy s radiochemickou separací (RPA) a protony vyvolané emise paprsků gama (PIGE) pro určování fluoru v biologických a ekologických materiálech jsou navzájem srovnány s ohledem na detekční limity, jaderné a jiné interference.
    Trvalý link: http://hdl.handle.net/11104/0000082

     
     
Počet záznamů: 1  