Number of the records: 1  

Characterization of the radiation environment by liulin-type spectrometers

  1. 1.
    0360649 - ÚJF 2012 RIV GB eng J - Journal Article
    Dachev, T. - Spurný, František - Ploc, Ondřej
    Characterization of the radiation environment by liulin-type spectrometers.
    Radiation Protection Dosimetry. Roč. 144, 1-4 (2011), s. 680-683. ISSN 0144-8420. E-ISSN 1742-3406
    Institutional research plan: CEZ:AV0Z10480505
    Keywords : instrument * dosimetry * models * Chandrayaan-1
    Subject RIV: BG - Nuclear, Atomic and Molecular Physics, Colliders
    Impact factor: 0.822, year: 2011

    Liulin-type spectrometers can characterise the type of predominant particles and their energy in the radiation environment. The results from calibrations and space and aircraft experiments revealed that the most informative is by the shape of the deposited energy spectrum. Spectra generated by galactic cosmic rays (GCR) protons and their secondaries look like straight lines in the coordinates deposited energy/deposited per channel dose rate. The position of the maximum of the deposited energy spectra depends on the incident energy of the incoming protons. Spectra generated by relativistic electrons in the outer radiation belt have a maximum in the first channels. For higher energy depositions, these spectra are similar to the GCR spectra. All types of spectra have a knee close to 6.3 MeV of deposited energy, which corresponds to the stopping energy of protons in the detector.
    Permanent Link: http://hdl.handle.net/11104/0198149

     
     