Počet záznamů: 1

Surface Plasmon Resonance Biosensor for Determination of Tetrodotoxin: Prevalidation Study

  1. 1.
    0367829 - URE-Y 2012 RIV US eng J - Článek v odborném periodiku
    Vaisocherová, Hana - Taylor, A. D. - Jiang, S. - Hegnerová, Kateřina - Vala, Milan - Homola, Jiří - Yakes, B. - Deeds, J. - DeGrasse, S.
    Surface Plasmon Resonance Biosensor for Determination of Tetrodotoxin: Prevalidation Study.
    Journal of Aoac International. Roč. 94, č. 2 (2011), s. 596-604 ISSN 1060-3271
    Grant CEP: GA AV ČR KAN200670701
    Výzkumný záměr: CEZ:AV0Z20670512
    Klíčová slova: toxin detection * biosensor * validtion
    Kód oboru RIV: JA - Elektronika a optoelektronika, elektrotechnika
    Impakt faktor: 1.199, rok: 2011

    A label-free surface plasmon resonance biosensor method was applied to determine tetrodotoxin (TTX) in pufferfish matrixes using an antibody inhibition assay format. A prevalidation study was conducted to demonstrate the assay performance characteristics, such as selectivity, LOD, LOQ, repeatability, reproducibility, and accuracy.The developed method was demonstrated to be capable of detecting TTX in pufferfish matrix standard samples in a broad concentration range (2-9000 ng/mL) with an LOD of 1.5 ng/mL.