Acta Chir Orthop Traumatol Cech. 2012; 79(3):213-221 | DOI: 10.55095/achot2012/031

Current Knowledge on the Effect of Technology and Sterilization on the Structure, Properties and Longevity of UHMWPE in Total Joint ReplacementCurrent concepts review

D. POKORNÝ1,*, M. ŠLOUF2, P. FULÍN1
1 I. Ortopedická klinika 1. LF UK Praha, FN Motol, Praha
2 Ústav makromolekulární chemie AV ČR, v.v.i., Praha

Ultrahigh molecular weight polyethylene (UHMWPE) is the most frequently used bearing surface in currently used total joint replacements (TJR). According to the literature available, UHMWPE is the best polymer material, in terms of biocompatibility, mechanical properties and wear resistance, for this application. In spite of this fact, UHMWPE wear (i.e., release of microscopic particles from the polymer surface) remains one of the main reasons of TJR failures. Consequently, the wear of UHMWPE is a subject of intensive study by both materials scientists and orthopaedic surgeons.
The structure and properties of UHMWPE strongly depend on the way of processing and post-processing modifications. The processing includes polymer resin preparation (microparticles about 100 μm in size) and resin consolidation (forming bulk material). Post-processing modifications aim at increasing wear resistance and oxidation stability which are regarded as major factors involved in TJR failure. In order to maintain high purity materials for medical application, it is not allowed to use additional chemicals during the modification processes. The only exception is the use of vitamin E, a natural stabilizer and antioxidant. Considering all the above mentioned facts, the modifications can be based on (I) ionizing radiation such as gamma rays or accelerated electrons, (II) thermal modification, (III) additional stabilization with vitamin E, and (IV) sterilization. According to the modifications, we usually differentiate three generations of UHMWPE.
The 1st generation UHMWPE is not modified except for obligatory sterilization. The sterilization procedures based on chemical procedures (formaldehyde vapours, hot water) have lately been forbidden, abandoned and replaced by gamma-irradiation with doses of 25-45 kGy. In the course of time, sterilization by means of gamma-irradiation showed to be unsuitable due to oxidative degradation of UHMWPE, which resulted in lower wear resistance, worse material properties and a decrease in longevity of joint replacements.
The 2nd generation UHMWPE or highly cross-linked UHMWPE is material treated with ionizing radiation (gamma-rays or accelerated electrons) at higher doses (usually 50-100 kGy). The irradiation is followed by thermal treatment in order to eliminate residual radicals and limit oxidative degradation that occurs in the 1st generation UHMWPE types. Finally, the material is sterilized usually by modern procedures using ethylene oxide or gas plasma neither of which causes oxidative degradation (as opposed to gamma-irradiation sterilization).
The 3rd generation UHMWPE is basically the same as that of the 2nd generation. The main difference consists in slightly better mechanical properties (strength, fatigue) and further stabilization to oxidative degradation. This is achieved by special procedures, some of which are either controversial (sequential irradiation) or well-accepted by a great majority of the orthopaedic community (vitamin E stabilization). Only the future will show whether the difference between the 2nd and 3rd generations of UHMWPE will play any role in increasing TJR longevity and therefore the quality of its performance. The final objective of all of UHMWPE modifications is, as already mentioned, to increase wear resistance, oxidative resistance and mechanical performance of the polymer. These changes, especially an increase in wear resistance, have been confirmed to prolong the longevity of total joint replacements based on UHMWPE.

Keywords: polyethylene, UHMWPE, cross-linking, arthroplasty, total hip arthroplasty, oxidative degradation, vitamin E, sequential irradiation, total joint replacements

Published: June 1, 2012  Show citation

ACS AIP APA ASA Harvard Chicago IEEE ISO690 MLA NLM Turabian Vancouver
POKORNÝ D, ŠLOUF M, FULÍN P. Current Knowledge on the Effect of Technology and Sterilization on the Structure, Properties and Longevity of UHMWPE in Total Joint Replacement. Acta Chir Orthop Traumatol Cech.. 2012;79(3):213-221. doi: 10.55095/achot2012/031. PubMed PMID: 22840952.
Download citation

References

  1. CHARNLEY, J.: Arthroplasty of the hip: a new operation. Lancet, 1129-1132, 1961. Go to original source... Go to PubMed...
  2. CHARNLEY, J.: Low friction principle. In: CHARNLEY, J.: Low friction arthroplasty of the hip: theory and practice. Berlin, Springer-Verlag, 1979. Go to original source...
  3. CHIESA, R., TANZI, M. C., ALFONSI, S., PARACCHINI, L., MOSCATELLI, M., CIGADA, A.: Enhanced wear performance of highly crosslinked UHMWPE for artificial joints. J. Biomed. Mater. Res., 50: 381-387, 2000. Go to original source...
  4. COLLIER, J. P., SPERLING, D. K., CURRIER, J. H., SUTULA, L. C., SAUM, K. A., MAYOR, M. B.: Impact of gammasterilization on clinical performance of polyethylene in the knee. J. Arthropl., 11: 377-389, 1996. Go to original source... Go to PubMed...
  5. COSTA, L., BRACCO, P., BRACH DEL PREVER, E. M., KURTZ, S. M., GALLINARO, P.: Oxidation and oxidation potential in contemporary packaging for polyethylene total joint replacement components. J. Biomed. Mater. Res. B Appl. Biomater., 78: 20-26, 2006. Go to original source... Go to PubMed...
  6. COSTA, L., BRACH, E. M., BRACCO, P., LUDA, M. P.: Stabilization of UHMWPE with Vitamin E. UHMWPE for arthroplasty: degradation, stabilisation and crosslinking. Torino. Italy. 2005.
  7. COSTA, L., LUDA, M. P., TROSSARELLI, L., BRACH DEL PREVER, E. M., CROVA, M., GALLINARO, P.: Oxidation in orthopaedic UHMWPE sterilized by gamma-radiation and ethylen oxide. Biomaterials, 19: 659-668, 1998. Go to original source... Go to PubMed...
  8. DOLEŽAL, B., ADAMÍROVÁ, L.: Způsob zdravotně nezávadné stabilizace polyolefinu proti termooxidační a fotooxidační degradaci. Patent ČSSR č. 221403, 1982.
  9. GALLO, J., ŠLOUF, M., GOODMAN, S. B.: The relationship of polyethylene wear to particle size, distribution, and number: A possible factor explaining the risk of osteolysis after hip arthroplasty. J. Biomed. Mater. Res. Part B: Appl. Biomater., in press; DOI: 10.1002/jbm.b.31638.
  10. KURTZ, S. M.: A primer on UHMWPE. In: Kurtz, S. M.(eds): UHMWPE Biomaterials handbook, second edition. London, Elsevier - AcademicPress, 2009, 1-6. Go to original source...
  11. KURTZ, S. M.:The clinical performance of UHMWPE in hip replacements. In: Kurtz, S. M.(eds): UHMWPE Biomaterials handbook, second edition. London, Elsevier - AcademicPress, 2009, 43-54. Go to original source...
  12. KURTZ, S. M.: Vitamin-E-Blended UHMWPE Biomaterials. In: Kurtz, S. M.(eds): UHMWPE Biomaterials handbook, second edition. London, Elsevier - AcademicPress, 2009, 237-246. Go to original source...
  13. LANDY, M. M., WALKER, P.S.: Wearof Ultra-High-Molecular-Weight Polyethylene Components of 90 retrieved knee prostheses. J. Arthropl., 3-Suppl: 73-85, 1988. Go to original source... Go to PubMed...
  14. MATTHEWS, J. B., BESONG, A. A., GREEN, T. R., STONE, M. H., WROBELSKI, B. M., FISHER, J., INGHAM, E.: Evaluation of the response of primary human peripheral blood mononuclear phagocytes to challenge with in vitro generated clinically relevant UHMWPE particles of known size and dose. J. Biomed. Mater. Res.,52: 296-307, 2000. Go to original source...
  15. McKELLOP, H. A., CAMPBELL, P., PARK, S. H., SCHMALZRIED, T. P., GRIGORIS, P., AMSCHUTZ, H. C., SARMIENTO, A.: The origin of submicronpolyethylene wear debris in total hip arthroplasty. Clin. Orthop., 311: 3-20, 1995.
  16. McKELLOP, H. A., SHEN, F. W., LU, B., CAMPBELL, P., SALOVEY, R.: Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene. A hip-simulator study. J. Bone Jt Surg., 82-A: 1708-1725, 2000. Go to original source... Go to PubMed...
  17. OONISHI, H., KADOYA, Y.: Wear of high-dose gamma-irradiated polyethylene in total hip replacements. J. Orthop. Sci., 5: 223-228, 2000. Go to original source... Go to PubMed...
  18. OONISHI, H., KADOYA, Y., MASUDA, S.: Gamma-irradiated cross-linked polyethylene in total hip replacements - Analysis of retrieved sockets after long-term implantation. J. Biomed. Mater. Res., 58: 167-171, 2001. Go to original source...
  19. ORAL, E.: Highly crosslinked UHMWPE doped with vitamin E. In: Kurtz, S. M.(eds): UHMWPE Biomaterials handbook, second edition. London, Elsevier - AcademicPress, 2009, 221-233. Go to original source...
  20. POKORNÝ, D., ŠLOUF, M., FULÍN, P., DANIEL, M., SOSNA, A.: Otěrové částice a jejich lokální a celkový negativní vliv. In: Landor, I., Vavřík, P., Gallo, J., Sosna, A.: Revizní operace totální náhrady kyčelního kloubu. ISBN 978-80-7345-254-4, v tisku.
  21. PREMNATH, V., HARRIS, W. H., JASTY, M., MERRILL, E. W.: Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Ultra High Molecular Weight Poly Ethylene. Biomaterials, 17: 1741-1753, 1996. Go to original source... Go to PubMed...
  22. REEVES, E. A., BARTON, D. C., FITZPATRICK, D., FISHER, J.: Comparison of gas plasma and gamma irradiationin air sterilization on the delamination wear of the ultra-high molecular weight polyethylene used in knee replacements. Proc. Inst. Mech. Eng., Part H, 214: 249-255, 2000. Go to original source... Go to PubMed...
  23. ROTH, A., SANDER, K., LAYHER, F., BABISCH, J., VENBROCKS, R.: In vivo measurement of polyethylene wear in cementless total hip arthroplasty. Acta Chir. orthop. Traum. čech., 77: 13-17, 2010. Go to original source...
  24. SCHMALZRIED, T. P., JASTY, M., HARRIS, W. H.: Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space. J. Bone Jt Surg., 74-A: 849-863, 1992. Go to original source...
  25. SCHMALZRIED, T. P., JASTY, M., ROSENBERG, A., HARRIS, W. H.: Polyethylene wear debris and tissue reactions in knee as compared to hip replacement prostheses. J. Appl. Biomater., 5: 185-190, 1992. Go to original source... Go to PubMed...
  26. SHANBHAG, A.S., JACOBS, J.J., BLACK, J., GALANTE, J.O., GLANT, T. T.: Macrophage/ particle interactions: effect of size, composition and surface area. J. Biomed. Mater. Res., 28: 81-90, 1994. Go to original source... Go to PubMed...
  27. SOSNA, A., RADONSKÝ, T., POKORNÝ, D., VEIGL, D., HORÁK, Z., JAHODA, D.: Polyetylenová choroba. Acta Chir. orthop. Traum. čech., 70: 6-16, 2003.
  28. SUTULA, L. C., COLLIER, J. P., SAUM, K. A., CURRIER, B. H., CURRIER, J. H., SANFORD, W. M., MAYOR, M. B., WOODING, R. E., SPERLING, D. K., WILLIAMS, I. R.: The Otto Aufranc Award. Impact of gammasterilization on clinical performance of polyethylene in the hip. Clin. Orthop., 319: 28-40, 1995. Go to original source...
  29. ŠLOUF, M., BALDRIAN, J., DYBAL, J., PAWLOVÁ, E., LAPČÍKOVÁ, M., KOTEK, J., MIKEŠOVÁ, J., PILAŘ, J.: Výroba a charakterizace síťovaného UHMWPE pro kloubní náhrady II. Research report. Edice Makro: T-743, 2009.
  30. ŠLOUF, M., MIKEŠOVÁ, J., FENCL, J., SYNKOVÁ, H., BALDRIAN, J., HORÁK, Z.: Impact of dose-rate on rheology, structure and wear of irradiated UHMWPE. J. Macromol. Sci. Phys., 48: 587-603, 2009. Go to original source...
  31. ŠLOUF, M., SYNKOVÁ, H., BALDRIAN, J., MAREK, A., KOVÁŘOVÁ, J., SCHMIDT, P., DORSCHNER, H., STEPHAN, M., GOHS, U.: Structural changes of UHMWPE after e-beam irradiation and thermal treatment. J. Biomed. Mater. Res., Part B - Appl. Biomater. 85B: 240-251, 2008. Go to original source... Go to PubMed...
  32. WILLIAMS, I. R., MAYOR, M. B., COLLIER, J. P.: The impact of sterilization method on wear in knee arthroplasty. Clin. Orthop., 356, 170-180, 1998. Go to original source... Go to PubMed...
  33. WROBLEWSKI, B. M., SINEY, P. D., DOWSON, D., COLLINS, S. N.: Prospective clinical and joints simulator studies of a new total hip arthroplasty using aluminaceramic heads and cross-linked polyethylene cups. J. Bone Jt Surg., 78-B: 280-285, 1996. Go to original source...