Acta Chir Orthop Traumatol Cech. 2011; 78(1):49-55 | DOI: 10.55095/achot2011/008

Comparison of in Vivo Characteristics of Polyethylene Wear Particles Produced by a Metal and a Ceramic Femoral Component in Total Knee ReplacementOriginal papers

D. VEIGL1, P. VAVŘÍK1,*, D. POKORNÝ1, M. ŠLOUF2, E. PAVLOVA2, I. LANDOR1, P. Fulín1
1 I. ortopedická klinika 1. LF UK a FN Motol, Praha
2 Ústav makromolekulární chemie AV ČR, Praha

PURPOSE OF THE STUDY:
The aim of the study was to evaluate in vivo and compare, in terms of the quality and number of ultra high-molecular polyethylene (UHMWPE) wear particles, total knee replacements of identical construction differing only in the material used for femoral component production, i.e., CoCrMo alloy or ZrO2 ceramics.

MATERIAL AND METHODS:
Samples of peri-prosthetic granuloma tissue were collected in two patients with total knee replacement suffering from implant migration, who were matched in relevant characteristics. The primary knee replacement in Patient 1 with a CoCrMo femoral component was done 7.2 years and in Patient 2 with a ZrO2 implant 6.8 years before this assessment. The polyethylene wear-induced granuloma was analysed by the MORF method enabling us to assess the shape and size of wear debris and the IRc method for assessment of particle concentration.

RESULTS:
In the granuloma tissue samples of Patient 1, on the average, particles were 0.30 mm in size and their relative volume was 0.19. In the Patient 2 tissue samples, the average size of particles was 0.33 mm and their relative volume was 0.26. There was no significant difference in either particle morphology or their concentration in the granuloma tissue between the two patients.

DISCUSSION:
One of the options of how to reduce the production of polyethylene wear particles is to improve the tribological properties of contacting surfaces in total knee replacement by substituting a cobalt-chrome femoral component with a zirconia ceramic femoral component. The previous in vitro testing carried out with a mechanical simulator under conditions approaching a real weight-bearing in the human body did show a nearly three-fold decrease in the number of UHMWPE wear particles in zirconia components. The evaluation of granuloma tissue induced by the activity of a real prosthetic joint for nearly seven years, however, did not reveal any great difference in either quality or quantity of polyethylene debris between the two replacements. The difference of surface roughness between CoCrMo (Ra = 0.05) and ZrO2 (Ra = 0.02) components did not play any role in in vivo conditions.

CONCLUSIONS:
In accordance with a previous clinical study, this evaluation of the quality and quantity of UHMWPE wear particles produced by a ceramic femoral component in vivo failed to demonstrate any advantage of zirconia ceramic components over the cobalt-chrome femoral components so far used.

Keywords: total knee replacement, zirconia ceramic femoral component, comparative study, polyethylene wear, in vivo measurement

Published: February 1, 2011  Show citation

ACS AIP APA ASA Harvard Chicago IEEE ISO690 MLA NLM Turabian Vancouver
VEIGL D, VAVŘÍK P, POKORNÝ D, ŠLOUF M, PAVLOVA E, LANDOR I, Fulín P. Comparison of in Vivo Characteristics of Polyethylene Wear Particles Produced by a Metal and a Ceramic Femoral Component in Total Knee Replacement. Acta Chir Orthop Traumatol Cech.. 2011;78(1):49-55. doi: 10.55095/achot2011/008. PubMed PMID: 21375966.
Download citation

References

  1. BARTEL, D. L., BICKNELL, V. L., WRIGHT, T. M.: The effect of conformity thickness and material on stresses in ultra-high molecular weight components for total joint replacement. J. Bone Jt Surg. A-68: 1041-1051, 1986. Go to original source...
  2. BEK, J., VAVŘÍK, P., LANDOR, I.: Hodnocení přežití náhrady kolenního kloubu Walter-Motorlet - dlouhodobé výsledky. Acta Chir. orthop. Traum. čech., 72: 221-227, 2005.
  3. BEN-NISSAN, B., DOYLE, CH: The Use of Ceramics in Knee Replacement. TribOS Newsletter, 8: 11-13, 2009.
  4. DONÁT, A., KONVIČKOVÁ, S.: Biomechanics of the Ceramic Knee Replacement. Med. Biol. Eng. Comput., 37: 314-316, 1999. Go to original source... Go to PubMed...
  5. DANĚK, J., DENK, F., HLAVÁČEK, I., NEDOMA, J., STEHLÍK, J.,VAVŘÍK, P.: On the Stress-Strain Analysis of the Knee Replacement. In: Lagana, A., Gavrilova, M. L., Kumar, V., Mun, Y., Tan, C. J. K., Gervasi, O. (eds.): Computational Science and Its Applications - ICCSA 2004 Lecture Notes in Computer Science. Vol. 3044. Heidelberg, Springer Verlag 2004, 456-466. Go to original source...
  6. DANĚK, J., NEDOMA, J., HLAVÁČEK, I., VAVŘÍK, P., DENK, F.: Numerical modelling of the weight-bearing total knee joint replacement and usage in practice. Math. Comput. Simul., 76: 49-56, 2007. Go to original source...
  7. FULÍN, P., POKORNÝ, D., ŠLOUF, M., LAPČÍKOVÁ, M., PAVLOVA, E., ZOLOTAREVOVÁ E.: Metoda MORF pro sledování velikostí a tvarů otěrových mikro- a nanočástic UHMWPE v periprotetických tkáních. Acta Chir. orthop. Traum. čech., v tisku.
  8. GALLO, J., MRÁZEK, F., PETEREK, M: Podíl imunogenetických faktorů na vzniku a rozvoji a rozvoji periprotetické osteolýzy. Acta Chir. orthop. Traum. čech., 74: 247-252, 2007. Go to original source...
  9. GALLO, J., ZDAŘILOVÁ, A., RAJNOCHOVÁ SVOBODOVÁ, A., ULRICHOVÁ, J., RADOVÁ, L., SMIŽANSKÝ, M.: Výpotky asepticky selhávajících TEP kyčlí a kolen nejsou toxické pro osteoblasty. Acta Chir. orthop. Traum. čech., 77: 416-424, 2010. Go to original source...
  10. HAIDER, H.: Features and Difficulties of Knee Wear Simulation. TribOS Newsletter, 9: 18-21, 2009.
  11. KADDICK, CH.,WIMMER, M.: Polyethylene thickness- is it still relevant ? TribOS Newsletter, 8: 9-10, 2009.
  12. KAPLAN, E. L., MEIER, P.: Nonparametric Estimation from Incomplete Observations. J. Amer. Stat. Assoc., 53: 457-481, 1958. Go to original source...
  13. KUČERA, T., Urban, K., KARPAŠ, K., ŠPONER,P.: Omezení hybnosti kolenního kloubu po implantaci totální endoprotézy. Acta Chir. orthop. Traum. čech., 74: 326-331, 2007. Go to original source...
  14. LANGOVÁ, K., GALLO, J.: Je Kaplan-Meierova statistika nejvhodnější metodou k hodnocení přežívání výsledků v ortopedii? Acta Chir. orthop. Traum. čech., 77: 118-123, 2010. Go to original source...
  15. LAPČÍKOVÁ, M., SLOUF, M., DYBAL, J., ZOLOTAREROVA, E., ENTLICHER, G., POKORNÝ, D., GALLO, J., SOSNA, A.: Nanometer size wear debris generated from ultra high molecular weight polyethylene in vivo. Wear, 266: 349-355, 2009. Go to original source...
  16. MINODA, Y., KOBAYASHI, A., IWAKI, H., IWAKIRI, K., INORI, F., SUGAMA, R., IKEBUCHI, M., KADOYA, Y., TAKAOKA, K.: In Vivo Analysis of Polyethylene Wear Particles After Total Knee Arthroplasty: The Influence of Improved Materials and Designs. J. Bone Jt Surg., 91-A: 67-73, 2009. Go to original source... Go to PubMed...
  17. NAKAMURA, T., OONISHI, E., YASUDA, T., NAKAGAWA, Y.: A new knee prosthesis with bisurface femoral komponent made of zirconia ceramic (report 2). Key Eng. Mat., 254-256: 607-610, 2004. Go to original source...
  18. NAKAMURA, S., KOBAYASHI, M., ITO, H., NAKAMURA, K., UEO, T., NAKAMURA, T.: The Bi-Surface total knee arthroplasty: minimum 10-year follow-up study. Knee, 17: 274-278, 2010. Go to original source... Go to PubMed...
  19. NAUDIE, D. R., AMMEEN, D. J., ENGH, G. A., RORABECK, C. H.: Wear and Osteolysis Around Total Knee Arthroplasty. J. Amer. Acad. Orthop. Surf.,15: 53-64, 2007.
  20. NEDOMA, J., HLAVÁČEK, I., DANĚK, J., VAVŘÍK, P., STEHLÍK, J., DENK, F.: Some recent Results On a Domain Decomposition Method in Biomechanics of Human Joints. In: Kumar, V., Gavrilova, M. L., Tan, C. J. K. (eds.): Computational Science and Its Applications - ICCSA 2003 Lecture Notes in Computer Science. Vol. 2667: Heidelberg, Springer Verlag 2003, 587-600. Go to original source...
  21. NEDOMA, J., STEHLÍK, J., BARTOŠ, M., DENK, F., DŽUPA, V., FOUSEK, J., HLAVÁČEK, I., KLÉZL, Z., KVĚT, I.: Biomechanika lidského skeletu a umělých náhrad jeho částí. Praha, Nakl. Karolinum 2006.
  22. OKA, M., KUMAR, P., IKEUCHI, T. et al.: Low wear rate of UWHMPE against zirconia ceramic. Bioceramics, 5: 373-379, 1992.
  23. OONISHI, H., UENO, M., KIM, S.C., OONISHI, H., IWAMOTO, M., KYOMOTO, M.: Ceramic versus cobalt-chrome femoral components; wear of polyethylene insert in total knee prosthesis. J. Arthroplasty, 24: 374-382, 2009. Go to original source... Go to PubMed...
  24. OONOSHI, H., MURATA, N., SAITO, M., et al.: 3 to 18 Year Clinical Results of Total Knee Replacement with Ceramic Components. Key Eng. Mat., 192-195: 999-1002, 2001. Go to original source...
  25. PAYTEN, M. W., BEN-NISSAN, B.: Development of a modular ceramic knee prosthesis. In: Wise, D. L., Trantolo, D. J., Lewandrowski, K. U., Gresser, J. D., Cattaneo, M. V., Yaszemski, M. J. (eds): Biomaterials Engineering and Devices. Human applications. New Jersey, Humana Press 2000, 309-336. Go to original source...
  26. POKORNÝ, D., JAHODA, D., TOMAIDES, J., VAVŘÍK, P., LANDOR, I., SOSNA, A.: Sedmileté klinické výsledky systému totální náhrady kolenního kloubu SVL / Beznoska. Acta Chir. orthop. Traum. čech., 72: 277-281, 2005.
  27. POKORNÝ D., ŠLOUF M., DYBA, J., ZOLOTAREVOVÁ, E., VESELÝ, F., FULÍN, P., JAHODA. D.,VAVŘÍK, P., ENTLICHER, G., SOSNA, A: Nová metoda kvantifikace otěrových částic UHMWPE v okolí kloubních náhrad. Acta Chir. orthop. Traum. čech., 76: 374-381, 2009. Go to original source...
  28. POKORNÝ, D., ŠLOUF, M., VESELÝ, F., FULÍN, P., JAHODA, D., LANDOR, I., POPELKA, S., BĚLÁČEK, J., ZOLOTAREVOVA, E., SOSNA, A.: Distribuce otěrových částic UHMWPE v periprotetických tkáních u TEP kyčelního kloubu. Acta Chir. orthop. Traum. čech., 77: 87-92, 2010. Go to original source...
  29. ROBA, M., NAKA, M., GAUTIER, E., SPENCER, N.D., CROCKETT, R.: The adsorption and lubrication behaviorof synovial fluid proteins and glycoproteins on the bearing surface materials of hip replacements. Biomaterials, 30: 2072-2078, 2009. Go to original source... Go to PubMed...
  30. ROTH, A., SANDER, K., LAYHER, F., BABISCH, J., VENBROCKS, R.: In Vivo Measurement of Polyethylene Wear in Cementless Total Hip Arthroplasty. Acta Chir. orthop. Traum. čech., 77: 13-17, 2010. Go to original source...
  31. ROZKYDAL, Z., JANÍK, P., JANÍČEK, P., KUNOVSKÝ, R.: Revizní náhrada kolena po aseptickém uvolnění. Acta Chir. orthop. Traum. čech., 74: 5-13, 2007. Go to original source...
  32. RYBKA, V., VAVŘÍK, P., LANDOR,I.: Assessment of the Walter-Motorlet cemented condylar knee replacement - 100 cases. J. Orthop. Rheum., 4: 71-81, 1991.
  33. SIDA, D., BLAISE. L., SIDA, V., VAVRIK, P.: Zirconia ceramic femoral components. Key Eng. Mat., 218: 581-584, 2002. Go to original source...
  34. ŠÍDA, V.: Zpráva o průběhu a výsledcích řešení projektu "Kondylární náhrada kolenního kloubu s aplikací keramiky" č. PPZ1/29/98. DIAS a.s., Turnov, 1998.
  35. SOSNA, A., RADONSKÝ, T., POKORNÝ, D., VEIGL, D., HORÁK, Z., JAHODA, D.: Polyetylenová choroba. Acta Chir. orthop. Traum. čech., 70: 6-16, 2003.
  36. SPECTOR, M., RIES, M. D., Bourne, R. B., SAUER, S. W., LONG, M., HUNTER, G.: Wear performance of ultra-high molecular weight polyethylen on oxidized zirkonium total knee femoral components. J. Bone Jt Surg., 83-A: 80-86, 2001. Go to original source... Go to PubMed...
  37. ŠLOUF, M., POKORNÝ, D., ENTLICHER, G., DYBAL, J., SYNKOVÁ, H., LAPČÍKOVÁ, M., FEJFARKOVÁ, Z., ŠPUNDOVÁ, M., VESELÝ, F., SOSNA, A.: Quantification of UHMWPE wear debris in periprosthetic tissues of hip arthroplasty: description of a new method based on IR and comparison with radiographic appearance. Wear, 265: 674-684, 2008. Go to original source...
  38. TAYLOR, T.: Computational Wear Testing of TKR. TribOS Newsletter, 9: 14-17, 2009. Go to original source...
  39. TONI, A.: Realtionship Between Obesity and Early failure of Total Knee Prostheses. TribOS Newsletter, 8: 7-8, 2009.
  40. UENO, M., APGAR, M., SARIN, V.: Mechanical stress analysis and burst testing of a zirconia femoral component for total knee arthroplasty. Key Eng. Mat., 218-220: 573-576, 2002. Go to original source...
  41. VAVŘÍK, P., LANDOR, L: Revision TKA in the cases with severe polyethylene wear. Sborník přednášek: X. Národní kongres ČSOT s mezinárodní účastí. 18. 5. 2006, Praha.
  42. VAVŘÍK, P., LANDOR, I., DENK, F.: Klinické zhodnocení použití keramické femorální komponenty v konstrukci náhrady kolenního kloubu. Acta Chir. orthop. Traum. čech., 75: 436-442, 2008. Go to original source...
  43. VAVŘÍK P., LANDOR, I., TOMAIDES, J., POPELKA, S.: Střednědobé výsledky u náhrad kolenního kloubu Medin Modular. Acta Chir. orthop. Traum. čech., 76: 30-34, 2009. Go to original source...
  44. VAVŘÍK, P., POPELKA, S., VÍRAVA, J.: Ojedinělý případ oboustranného výskytu rozsáhlého polyetylénového granulomu po náhradě kolena. Acta Chir. orthop. Traum. čech., 68: 320-323, 2001.
  45. VAVŘÍK, P. VEIGL, D.: Uvolnění a periprotetická osteolýza kloubních náhrad. Čes. Revmatol., 10: 157-164, 2002.
  46. VEIGL, D., NIEDERLOVÁ, J., KRYŠTŮFKOVÁ, O.: Periprosthetic osteolysis and its association with the molecule RANKL expression. Physiol. Res., 56: 455-62, 2007. Go to original source... Go to PubMed...
  47. VESELÝ, F., NOVOTNÝ, K., SOSNA, A.: Záměna masivního heterotopického otěrového granulomu s aneuryzmatem arteria iliaca externa. Acta Chir. orthop. Traum. čech., 76: 509-512, 2009. Go to original source...
  48. WALKER, P.: Questions & Answers: Wear in Knees. TribOS Newsletter, 3: 10-11, 2008.
  49. ZOLOTAREVOVA, E., FEJFARKOVÁ, Z., ENTLICHER, G., LAPČÍKOVÁ, M., SLOUF, M., POKORNÝ, D., SOSNA, A.: Can centrifugation affect the morphology of polyethylene wear debris? Wear, 265: 1914-1917, 2008. Go to original source...